Open official notice

Cardinal health presource packs IV start kit- intended for IV access device catalog #: 01-5798b

Official notice

1 Mar 2023

RecallScope risk estimate:CriticalFDA

Cardinal Health Presource Packs IV START KIT- Intended for IV access device CATALOG #: 01-5798B

What to do now

Stop using the affected device and follow the corrective action in the official FDA notice. Contact your healthcare provider or the manufacturer for next steps.

Units affected

4,560

Recall number

Z-1157-2023

Official notice

2023-03-01

The hazard

Presource Kits contains the ICU MicroClave Clear Connector due to a manufacturing defect there is a gap between the connectors top and bottom housings which may lead to fluid leaks and loss of fluid path integrity

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Lots

816643

Other

UDI-DI 50195594577153

Event Ids

91560

Recall Numbers

Z-1157-2023

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Cardinal Health Presource Packs IV START KIT- Intended for IV access device CATALOG #: 01-5798B

Units affected

4,560

Hazard

Presource Kits contains the ICU MicroClave Clear Connector due to a manufacturing defect there is a gap between the connectors top and bottom housings which may lead to fluid leaks and loss of fluid path integrity

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Company & contacts

Manufacturer

Cardinal Health 200, LLC

Phone

Not provided

Brand: Cardinal Health 200. Contact: Cardinal Health 200, LLC.

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-1157-2023

Event ID

91560

Units affected

4,560

Status

active

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 1 Mar 2023. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button