Open official notice

Atlantis(r) anterior cervical plate system variable angle screw module, ref 176-503, qty: 1 ea, Rx only, medtronic sofamore danek usa

Official notice

17 Oct 2012

RecallScope risk estimate:MediumFDA

An official FDA notice concerns Atlantis(r) anterior cervical plate system variable angle screw module, ref 176-503, qty: 1 ea, Rx only, medtronic sofamore danek usa. Lines on the screw gauge may not placed on the caddy properly which could result in improper measurement of screws. An incorrect gauge could measure a screw to be approximately 1 mm shorter than its actual length. Published 2012-10-17.

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

1,047

Recall number

Z-0036-2013

Official notice

2012-10-17

The hazard

Lines on the screw gauge may not placed on the caddy properly which could result in improper measurement of screws. An incorrect gauge could measure a screw to be approximately 1 mm shorter than its actual length.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

63157

Recall Numbers

Z-0036-2013

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

ATLANTIS(R) Anterior Cervical Plate System VARIABLE ANGLE SCREW MODULE, REF 176-503, QTY: 1 EA, Rx only, Medtronic Sofamore Danek USA, Inc., 1800 Pyramid Place, Memphis, Tennessee 38132. Orthopedic. This instrument is a precision device which may incorporate a measuring function and has uses as described on the label. Unless labeled for single use, this instrument may be re-used.

Units affected

1,047

Hazard

Lines on the screw gauge may not placed on the caddy properly which could result in improper measurement of screws. An incorrect gauge could measure a screw to be approximately 1 mm shorter than its actual length.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Affected area 2

AU · Australia

Affected area 3

CA · Canada

Affected area 4

HK · Hong Kong

Affected area 5

IN · India

Affected area 6

JP · Japan

Affected area 7

KR · South Korea

Affected area 8

MX · Mexico

Affected area 9

NL · Netherlands

Affected area 10

SG · Singapore

Company & contacts

Manufacturer

Medtronic Sofamor Danek USA Inc

Phone

Not provided

Brand: Medtronic Sofamor Danek. Contact: Medtronic Sofamor Danek USA Inc.

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-0036-2013

Event ID

63157

Units affected

1,047

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 17 Oct 2012. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button