A mobile, electro-hydraulically operated, fluoroscopic C-Arm compatible general surgical table. Product Usage: Product provides a flexible

Official notice

26 Sept 2012

RecallScope risk estimate:MediumFDA

An official FDA notice concerns A mobile, electro-hydraulically operated, fluoroscopic C-Arm compatible general surgical table. Product Usage: Product provides a flexible. Customers may be storing objects on the base and/or around the column of the surgical table. This prohibited practice and misuse known by the customers can damage the override switch assembly located at the top of the column cover. Depending upon the nature and severity of the damage to the electronics of the table, it is possible that the table could move without prompting by an operator . Published 2012-09-26.

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

11,044

Recall number

Z-2402-2012

Official notice

2012-09-26

The hazard

Customers may be storing objects on the base and/or around the column of the surgical table. This prohibited practice and misuse known by the customers can damage the override switch assembly located at the top of the column cover. Depending upon the nature and severity of the damage to the electronics of the table, it is possible that the table could move without prompting by an operator .

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

54705

Recall Numbers

Z-2402-2012

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

A mobile, electro-hydraulically operated, fluoroscopic/C-Arm compatible general surgical table. Product Usage: Product provides a flexible, easy-to-use, articulated posturing of a patient for either general, I.A., or specialized surgical procedures. There are movable components and accessories intended for use during diagnostic examinations or surgical procedures to support and position a patient.

Units affected

11,044

Hazard

Customers may be storing objects on the base and/or around the column of the surgical table. This prohibited practice and misuse known by the customers can damage the override switch assembly located at the top of the column cover. Depending upon the nature and severity of the damage to the electronics of the table, it is possible that the table could move without prompting by an operator .

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Affected area 2

AR · Argentina

Affected area 3

AU · Australia

Affected area 4

BR · Brazil

Affected area 5

CA · Canada

Affected area 6

CN · China

Affected area 7

CO · Colombia

Affected area 8

CR · Costa Rica

Affected area 9

DE · Germany

Affected area 10

DO · Dominican Republic

Affected area 11

EG · Egypt

Affected area 12

ES · Spain

Affected area 13

GB · United Kingdom

Affected area 14

GR · Greece

Affected area 15

HK · Hong Kong

Affected area 16

IL · Israel

Affected area 17

IN · India

Affected area 18

IT · Italy

Affected area 19

JO · Jordan

Affected area 20

KR · South Korea

Affected area 21

KW · Kuwait

Affected area 22

MX · Mexico

Affected area 23

MY · Malaysia

Affected area 24

NZ · New Zealand

Affected area 25

PE · Peru

Affected area 26

PL · Poland

Affected area 27

PT · Portugal

Affected area 28

RU · Russia

Affected area 29

SA · Saudi Arabia

Affected area 30

SG · Singapore

Affected area 31

TH · Thailand

Affected area 32

TR · Turkey

Affected area 33

TW · Taiwan

Affected area 34

ZA · South Africa

Company & contacts

Brand

Manufacturer

Steris Corporation

Phone

Not provided

Brand: Steris. Contact: Steris Corporation.

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-2402-2012

Event ID

54705

Units affected

11,044

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 26 Sept 2012. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button