8000-021-002, Scopis¿ ENT Software with TGS¿ with Sofware Version: NOVA 3.6.0 RC16. Scopis is a next-generation solution for navigating

Official notice

29 May 2024

RecallScope risk estimate:MediumFDA

An official FDA notice concerns 8000-021-002, Scopis¿ ENT Software with TGS¿ with Sofware Version: NOVA 3.6.0 RC16. Scopis is a next-generation solution for navigating. Unreleased software was installed on customer systems resulting in the visual feedback on the screen to show the points to be off from the physical reference point of the pointer or suction. Published 2024-05-29.

What to do now

Stop using the affected device and follow the corrective action in the official FDA notice. Contact your healthcare provider or the manufacturer for next steps.

Units affected

705

Recall number

Z-1885-2024

Official notice

2024-05-29

The hazard

Unreleased software was installed on customer systems resulting in the visual feedback on the screen to show the points to be off from the physical reference point of the pointer or suction.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

94456

Recall Numbers

Z-1885-2024

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

8000-021-002, Scopis¿ ENT Software with TGS¿ with Sofware Version: NOVA 3.6.0 RC16. Scopis is a next-generation solution for navigating functional endoscopic sinus surgery (FESS) offering surgeons highly advanced image guidance and visualization capabilities in a single system. The Stryker Electromagnetic Navigation Unit (8000-010-003) powered by Scopis ENT Software with TGS (8000-021-002), allows the analysis and identification of sinus cells in the complex patient anatomy and planning of the natural drainage pathways through the sinus cavity. During surgery, the planned pathways are overlaid in real-time onto the endoscopic image, providing a unique Scopis augmented reality technology. Guidance of endoscopic instruments may help perform a minimally invasive, accurate and selective surgery.

Units affected

705

Hazard

Unreleased software was installed on customer systems resulting in the visual feedback on the screen to show the points to be off from the physical reference point of the pointer or suction.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Company & contacts

Manufacturer

Stryker Leibinger GmbH & Co. KG

Phone

Not provided

Brand: Stryker Leibinger Kg. Contact: Stryker Leibinger GmbH & Co. KG.

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-1885-2024

Event ID

94456

Units affected

705

Status

active

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 29 May 2024. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button