Open official notice

Trumatch ct cut guide kit r- trumatch patient specific instruments are intended to be used as patient-specific surgical instrumentation to

Official notice

8 Sept 2021

RecallScope risk estimate:MediumFDA

An official FDA notice concerns Trumatch ct cut guide kit r- trumatch patient specific instruments are intended to be used as patient-specific surgical instrumentation to. Software: A coding error associated with the Fast3D Segmentation software. During the scanning process the three images required of the hip, knee and ankle may not be aligned, causing the anatomic landmark locations to potentially result in limb malalignment. Published 2021-09-08.

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

9

Recall number

Z-2399-2021

Official notice

2021-09-08

The hazard

Software: A coding error associated with the Fast3D Segmentation software. During the scanning process the three images required of the hip, knee and ankle may not be aligned, causing the anatomic landmark locations to potentially result in limb malalignment.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

88522

Recall Numbers

Z-2399-2021

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

TRUMATCH CT CUT GUIDE KIT R- TruMatch Patient Specific Instruments are intended to be used as patient-specific surgical instrumentation to assist in the positioning of a joint replacement component intra-operatively and in guiding the marking of bone before cutting Part Number: 420916

Units affected

9

Hazard

Software: A coding error associated with the Fast3D Segmentation software. During the scanning process the three images required of the hip, knee and ankle may not be aligned, causing the anatomic landmark locations to potentially result in limb malalignment.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States · CA

Location 2

US · United States · CO

Location 3

US · United States · FL

Location 4

US · United States · IL

Location 5

US · United States · IN

Location 6

US · United States · MI

Location 7

US · United States · MN

Location 8

US · United States · NY

Location 9

US · United States · TX

Location 10

US · United States · WA

Affected area 1

US · United States

Affected area 2

GB · United Kingdom

Affected area 3

PL · Poland

Company & contacts

Manufacturer

DePuy Orthopaedics, Inc.

Phone

Not provided

Brand: Depuy Orthopaedics. Contact: DePuy Orthopaedics, Inc..

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-2399-2021

Event ID

88522

Units affected

9

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 8 Sept 2021. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button