VITROS¿ Black Reference Slides CATALOG #: J02316 - Product Usage: The Black Reference Slide is a part tool used in the reflectometer

Official notice

5 May 2021

RecallScope risk estimate:CriticalFDA

An official FDA notice concerns VITROS¿ Black Reference Slides CATALOG #: J02316 - Product Usage: The Black Reference Slide is a part tool used in the reflectometer. Delayed results. Ortho confirmed that two lots of VITROS Black Reference Slides are incorrectly labelled as White Reference Slides. If the customer fails to notice both the product name or the unexpected color of the slides and tries to use Black Reference Slides as White Reference Slides, the Correction Factor test will fail - and the analyzer will not be able to produce results for any assays other than sodium, potassium, and chloride. If the customer doesn t have any White Reference Slides in reserve, they can t run the Correction Factor test correctly and get the analyzer running again until they get more White Reference Slides delivered. This could lead to a delay in results for all Vitros assays except for sodium, potassium, and chloride. Published 2021-05-05.

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

865

Recall number

Z-1489-2021

Official notice

2021-05-05

The hazard

Delayed results. Ortho confirmed that two lots of VITROS Black Reference Slides are incorrectly labelled as White Reference Slides. If the customer fails to notice both the product name or the unexpected color of the slides and tries to use Black Reference Slides as White Reference Slides, the Correction Factor test will fail - and the analyzer will not be able to produce results for any assays other than sodium, potassium, and chloride. If the customer doesn t have any White Reference Slides in reserve, they can t run the Correction Factor test correctly and get the analyzer running again until they get more White Reference Slides delivered. This could lead to a delay in results for all Vitros assays except for sodium, potassium, and chloride.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

87649

Recall Numbers

Z-1489-2021

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

VITROS¿ Black Reference Slides CATALOG #: J02316 - Product Usage: The Black Reference Slide is a part/ tool used in the reflectometer correction factor adjustment procedure.

Units affected

865

Hazard

Delayed results. Ortho confirmed that two lots of VITROS Black Reference Slides are incorrectly labelled as White Reference Slides. If the customer fails to notice both the product name or the unexpected color of the slides and tries to use Black Reference Slides as White Reference Slides, the Correction Factor test will fail - and the analyzer will not be able to produce results for any assays other than sodium, potassium, and chloride. If the customer doesn t have any White Reference Slides in reserve, they can t run the Correction Factor test correctly and get the analyzer running again until they get more White Reference Slides delivered. This could lead to a delay in results for all Vitros assays except for sodium, potassium, and chloride.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Affected area 2

AU · Australia

Affected area 3

BE · Belgium

Affected area 4

BR · Brazil

Affected area 5

CA · Canada

Affected area 6

CN · China

Affected area 7

CO · Colombia

Affected area 8

DE · Germany

Affected area 9

DK · Denmark

Affected area 10

ES · Spain

Affected area 11

FR · France

Affected area 12

GB · United Kingdom

Affected area 13

IN · India

Affected area 14

IT · Italy

Affected area 15

JP · Japan

Affected area 16

MX · Mexico

Affected area 17

NL · Netherlands

Affected area 18

NO · Norway

Affected area 19

PL · Poland

Affected area 20

PT · Portugal

Affected area 21

RU · Russia

Affected area 22

SE · Sweden

Affected area 23

SG · Singapore

Company & contacts

Manufacturer

Ortho-Clinical Diagnostics, INc.

Phone

Not provided

Brand: Ortho-clinical Diagnostics. Contact: Ortho-Clinical Diagnostics, INc..

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-1489-2021

Event ID

87649

Units affected

865

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 5 May 2021. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button