Open official notice

Portrait Core Services, a set of software services that enable the communication and interaction of the Portrait Mobile Monitoring Solution

Official notice

22 Jan 2025

RecallScope risk estimate:MediumFDA

An official FDA notice concerns Portrait Core Services, a set of software services that enable the communication and interaction of the Portrait Mobile Monitoring Solution. There is a software issue that can result in the loss of patient monitoring on the Portrait Central Viewer Application after 425 days of continuous run-time. Published 2025-01-22.

What to do now

Stop using the affected device and follow the corrective action in the official FDA notice. Contact your healthcare provider or the manufacturer for next steps.

Units affected

7

Recall number

Z-0918-2025

Official notice

2025-01-22

The hazard

There is a software issue that can result in the loss of patient monitoring on the Portrait Central Viewer Application after 425 days of continuous run-time.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Other

UDI-DI 00195278659446

Event Ids

96027

Recall Numbers

Z-0918-2025

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Portrait Core Services, a set of software services that enable the communication and interaction of the Portrait Mobile Monitoring Solution components and will integrate into existing healthcare facility infrastructure and clinical information systems, Model CSSXB

Units affected

7

Hazard

There is a software issue that can result in the loss of patient monitoring on the Portrait Central Viewer Application after 425 days of continuous run-time.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Affected area 2

AE · United Arab Emirates

Affected area 3

AU · Australia

Affected area 4

CN · China

Affected area 5

CZ · Czech Republic

Affected area 6

FR · France

Affected area 7

GB · United Kingdom

Affected area 8

IT · Italy

Affected area 9

NL · Netherlands

Affected area 10

SG · Singapore

Company & contacts

Manufacturer

GE Medical Systems, LLC

Phone

Not provided

Brand: GE Medical. Contact: GE Medical Systems, LLC.

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-0918-2025

Event ID

96027

Units affected

7

Status

active

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 22 Jan 2025. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button