Open official notice

MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 13 cm, Curved Extensions, IC Tray REF:8888101001HP

Official notice

23 Aug 2023

RecallScope risk estimate:CriticalFDA

MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 13 cm, Curved Extensions, IC Tray REF:8888101001HP

What to do now

Stop using the affected device and follow the corrective action in the official FDA notice. Contact your healthcare provider or the manufacturer for next steps.

Units affected

2,370

Recall number

Z-2367-2023

Official notice

2023-08-23

The hazard

Catheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or excessive MDX may dislodge from the catheter lead to full catheter obstruction resulting in delay to treatment and partial obstruction resulting in reduced flow, or particulate dislodgement that may result in delay to treatment, hemolysis, embolism/embolus or thrombosis/thrombus

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Lots

2133600127

Event Ids

92660

Recall Numbers

Z-2367-2023

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 13 cm, Curved Extensions, IC Tray REF:8888101001HP

Units affected

2,370

Hazard

Catheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or excessive MDX may dislodge from the catheter lead to full catheter obstruction resulting in delay to treatment and partial obstruction resulting in reduced flow, or particulate dislodgement that may result in delay to treatment, hemolysis, embolism/embolus or thrombosis/thrombus

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Affected area 2

AE · United Arab Emirates

Affected area 3

AR · Argentina

Affected area 4

AT · Austria

Affected area 5

BE · Belgium

Affected area 6

BR · Brazil

Affected area 7

CA · Canada

Affected area 8

CH · Switzerland

Affected area 9

CN · China

Affected area 10

CR · Costa Rica

Affected area 11

CZ · Czech Republic

Affected area 12

DE · Germany

Affected area 13

DK · Denmark

Affected area 14

DO · Dominican Republic

Affected area 15

ES · Spain

Affected area 16

FI · Finland

Affected area 17

FR · France

Affected area 18

GB · United Kingdom

Affected area 19

GR · Greece

Affected area 20

HU · Hungary

Affected area 21

IE · Ireland

Affected area 22

IN · India

Affected area 23

IT · Italy

Affected area 24

KW · Kuwait

Affected area 25

MX · Mexico

Affected area 26

MY · Malaysia

Affected area 27

NL · Netherlands

Affected area 28

NO · Norway

Affected area 29

PL · Poland

Affected area 30

PT · Portugal

Affected area 31

SA · Saudi Arabia

Affected area 32

SE · Sweden

Affected area 33

SG · Singapore

Affected area 34

SK · Slovakia

Affected area 35

TH · Thailand

Affected area 36

TR · Turkey

Affected area 37

ZA · South Africa

Company & contacts

Manufacturer

Covidien LP

Phone

Not provided

Brand: Covidien Lp. Contact: Covidien LP.

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class I

Recall details

Recall number

Z-2367-2023

Event ID

92660

Units affected

2,370

Status

active

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 23 Aug 2023. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button