Image available on the official notice.
Powered wheelchairs (joystick type), “昭和貿易: Nutron and others” (manufactured 2006–2013) - Repair
Official notice
6 Mar 2015
Last checked
27 Oct 2026
An official CAA Japan notice concerns Powered wheelchairs (joystick type), “昭和貿易: Nutron and others” (manufactured 2006–2013) - Repair. Cases of unexpected acceleration have occurred in the United States due to a defect involving twisting of wire connections in the joystick. Due to a defect involving twisting of wire connections in the joystick of these products, cases have occurred in the United States in which the powered wheelchair continues traveling at a speed below the programmed speed and then accelerates unexpectedly when the speed recovers. Published 2015-03-06.
What to do now
Replace the joystick and driver control free of charge.
Units affected
584
Official notice
2015-03-06
The hazard
The hazard
Cases of unexpected acceleration have occurred in the United States due to a defect involving twisting of wire connections in the joystick. Due to a defect involving twisting of wire connections in the joystick of these products, cases have occurred in the United States in which the powered wheelchair continues traveling at a speed below the programmed speed and then accelerates unexpectedly when the speed recovers.
How to check if you're affected
How to check if you're affected
Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.
Lots
Models
Checked manually this time?
Add your VIN to the app Watchlist to receive alerts for matching official notices.
Turn on automatic VIN checksAffected products
Affected products
Product 1
Units affected
Hazard
Where it was sold
Where it was sold
Retailer information is not currently available in RecallScope. Check the official notice for distribution details.
Location 1
Affected area 1
Company & contacts
Company & contacts
Manufacturer
Phone
Contact: ■お客様のお問合せ先. Phone 06-6441-5612, 03-5623-5396.
Source & status
Source & status
Recall details
Recall details
Event id
Units affected
Source tier
Documents, notices & source
See the official notice and source documents for full details.
View full official noticeWeekly recall digest