Open official notice

Powered wheelchairs (joystick type), “昭和貿易: Nutron and others” (manufactured 2006–2013) - Repair

Official notice

6 Mar 2015

Last checked

27 Oct 2026

RecallScope risk estimate:MediumCAA Japan

An official CAA Japan notice concerns Powered wheelchairs (joystick type), “昭和貿易: Nutron and others” (manufactured 2006–2013) - Repair. Cases of unexpected acceleration have occurred in the United States due to a defect involving twisting of wire connections in the joystick. Due to a defect involving twisting of wire connections in the joystick of these products, cases have occurred in the United States in which the powered wheelchair continues traveling at a speed below the programmed speed and then accelerates unexpectedly when the speed recovers. Published 2015-03-06.

What to do now

Replace the joystick and driver control free of charge.

Units affected

584

Official notice

2015-03-06

The hazard

Cases of unexpected acceleration have occurred in the United States due to a defect involving twisting of wire connections in the joystick. Due to a defect involving twisting of wire connections in the joystick of these products, cases have occurred in the United States in which the powered wheelchair continues traveling at a speed below the programmed speed and then accelerates unexpectedly when the speed recovers.

How to check if you're affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Lots

07KE003341

Models

07KE003341

Checked manually this time?

Add your VIN to the app Watchlist to receive alerts for matching official notices.

Turn on automatic VIN checks

Affected products

Product 1

電動車いす (ジョイスティック型) 昭和貿易株式会社取扱いの「Nutron、Ranger X、Torque SP、TDX-SI、TDX-SP、TDX-Spree、TDX-SR、TDX-3」

Units affected

584

Hazard

Cases of unexpected acceleration have occurred in the United States due to a defect involving twisting of wire connections in the joystick. Due to a defect involving twisting of wire connections in the joystick of these products, cases have occurred in the United States in which the powered wheelchair continues traveling at a speed below the programmed speed and then accelerates unexpectedly when the speed recovers.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

JP · Japan

Affected area 1

JP · Japan

Company & contacts

Manufacturer

■お客様のお問合せ先

Phone

06-6441-5612, 03-5623-5396

Contact: ■お客様のお問合せ先. Phone 06-6441-5612, 03-5623-5396.

Source & status

Source

Tier

official

Category

VEHICLES

Official classification

消費者庁 — 修理

Recall details

Event id

CAA-00000011760

Units affected

584

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 6 Mar 2015. The next one won't wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.