Image not available in RecallScope.
Volista standop surgical light, ard568811961 vlt600df aim stp; ard568821961 vlt600df aim stp ql; ard568811911 vlt600sf aim stp
Official notice
25 Jul 2018
An official FDA notice concerns Volista standop surgical light, ard568811961 vlt600df aim stp; ard568821961 vlt600df aim stp ql; ard568811911 vlt600sf aim stp. Certain VOLISTA StandOP Surgical Lights have been identified as having a potential light head detachment issue. Improper assembly of the bracket that connects the light head to the axle fork could result in the light head bracket breaking, causing the light head to detach from the arm and become suspended only from the cabling inside the arm. Use of an affected device may result in serious injury to patient and/or medical staff if the light head detaches and drops, impacting patient and/or medical staff during demonstration, service, or use. Published 2018-07-25.
What to do now
This recall has been terminated. See the official FDA notice for the latest status.
Units affected
151
Recall number
Z-2502-2018
Official notice
2018-07-25
The hazard
The hazard
Certain VOLISTA StandOP Surgical Lights have been identified as having a potential light head detachment issue. Improper assembly of the bracket that connects the light head to the axle fork could result in the light head bracket breaking, causing the light head to detach from the arm and become suspended only from the cabling inside the arm. Use of an affected device may result in serious injury to patient and/or medical staff if the light head detaches and drops, impacting patient and/or medical staff during demonstration, service, or use.
How to check if you’re affected
How to check if you’re affected
Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.
Event Ids
Recall Numbers
Checked manually this time?
Follow brands and keywords in the app to receive alerts for matching official notices.
Turn on recall alertsAffected products
Affected products
Product 1
Units affected
Hazard
Where it was sold
Where it was sold
Retailer information is not currently available in RecallScope. Check the official notice for distribution details.
Location 1
Affected area 1
Affected area 2
Affected area 3
Affected area 4
Affected area 5
Affected area 6
Affected area 7
Affected area 8
Affected area 9
Affected area 10
Affected area 11
Affected area 12
Affected area 13
Affected area 14
Affected area 15
Affected area 16
Affected area 17
Affected area 18
Affected area 19
Affected area 20
Affected area 21
Affected area 22
Affected area 23
Affected area 24
Affected area 25
Affected area 26
Affected area 27
Affected area 28
Affected area 29
Affected area 30
Affected area 31
Affected area 32
Affected area 33
Affected area 34
Affected area 35
Affected area 36
Affected area 37
Affected area 38
Affected area 39
Affected area 40
Affected area 41
Affected area 42
Affected area 43
Affected area 44
Affected area 45
Affected area 46
Affected area 47
Affected area 48
Affected area 49
Affected area 50
Company & contacts
Company & contacts
Brand: Maquet Sas. Contact: Maquet SAS.
Source & status
Source & status
Recall details
Recall details
Recall number
Event ID
Units affected
Status
Source tier
Documents, notices & source
See the official notice and source documents for full details.
View full official noticeWeekly recall digest
Get next week’s recalls before you need them.
One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.
The digest arrives once a week. For matching recall alerts after notices are processed, get the app.