Open official notice

Recall of batch 217251300 of the Trendhip Stem Std 12/14 B-BRAUN T2S hip implant, catalogue number GL-HETHS402, due to an incorrect implant size on the labelling, which could lead to dislocation and/or loosening.

Official notice

16 Oct 2018

Last checked

26 Oct 2026

RecallScope risk estimate:Not specifiedAEMPS

An official AEMPS notice concerns Recall of batch 217251300 of the Trendhip Stem Std 12/14 B-BRAUN T2S hip implant, catalogue number GL-HETHS402, due to an incorrect implant size on the labelling, which could lead to dislocation and/or loosening.. Recall of batch 217251300 of the Trendhip Stem Std 12/14 B-BRAUN T2S hip implant, catalogue number GL-HETHS402, due to an incorrect implant size on the labelling, which could lead to dislocation and/or loosening. Published 2018-10-16.

What to do now

The AEMPS, having consulted its network of experts in orthopaedic surgery and traumatology, considers the company's general measures appropriate. However, regarding follow-up of patients implanted with the affected product, it recommends clinical and radiological follow-up for at least one year to rule out unexpected subsidence of the stem.

Recall number

2018-530

Official notice

2018-10-16

The hazard

Recall of batch 217251300 of the Trendhip Stem Std 12/14 B-BRAUN T2S hip implant, catalogue number GL-HETHS402, due to an incorrect implant size on the labelling, which could lead to dislocation and/or loosening.

How to check if you're affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Other

PS/PRV/47262

Raw Text

217251300

Recall Numbers

2018-530

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Implante de cadera TRENDHIP STEM STANDARD 12/14 B-BRAUN T2S. Numero de catálogo: GL-HET HS402

Hazard

Recall of batch 217251300 of the Trendhip Stem Std 12/14 B-BRAUN T2S hip implant, catalogue number GL-HETHS402, due to an incorrect implant size on the labelling, which could lead to dislocation and/or loosening.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

ES · Spain

Affected area 1

ES · Spain

Source & status

Source

Tier

official

Category

HEALTH

Official classification

AEMPS — Alerta de productos sanitarios (tipo 1A)

Recall details

Recall number

2018-530

Event id

2018-530

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 16 Oct 2018. The next one won't wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.