• Market recall of the recomLine ANA/ENA IgG assay kit (lot LAE121701) with strip tubes LEAd 487/3 and LEAd 487/4, because it may contain test strips that are too short (10cm instead of 10,5cm) and produce false negative results.

Official notice

21 Jun 2018

Last checked

26 Oct 2026

RecallScope risk estimate:HighAEMPS

An official AEMPS notice concerns • Market recall of the recomLine ANA/ENA IgG assay kit (lot LAE121701) with strip tubes LEAd 487/3 and LEAd 487/4, because it may contain test strips that are too short (10cm instead of 10,5cm) and produce false negative results.. • Market recall of the recomLine ANA/ENA IgG assay kit (lot LAE121701) with strip tubes LEAd 487/3 and LEAd 487/4, because it may contain test strips that are too short (10cm instead of 10,5cm) and produce false negative results. • Review of results previously obtained using the affected strip tubes. Published 2018-06-21.

What to do now

Consumer instructions are not currently available in RecallScope. Check the official notice for what to do.

Recall number

2018-316

Official notice

2018-06-21

The hazard

• Market recall of the recomLine ANA/ENA IgG assay kit (lot LAE121701) with strip tubes LEAd 487/3 and LEAd 487/4, because it may contain test strips that are too short (10cm instead of 10,5cm) and produce false negative results. • Review of results previously obtained using the affected strip tubes.

How to check if you're affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Other

PS/CGG/45905

Raw Text

LAE121701 (con tubos de tiras LEAd 487/3 y LEAd 487/4)

Recall Numbers

2018-316

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Kit de ensayo recomLine ANA/ENA IgG (Ref. 6072)

Hazard

• Market recall of the recomLine ANA/ENA IgG assay kit (lot LAE121701) with strip tubes LEAd 487/3 and LEAd 487/4, because it may contain test strips that are too short (10cm instead of 10,5cm) and produce false negative results. • Review of results previously obtained using the affected strip tubes.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

ES · Spain

Affected area 1

ES · Spain

Source & status

Source

Tier

official

Category

HEALTH

Official classification

AEMPS — Alerta de vigilancia de productos sanitarios (tipo 2A)

Recall details

Recall number

2018-316

Event id

2018-316

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 21 Jun 2018. The next one won't wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button