Open official notice

Recall of lots 262401008E and 262508007E of ZARYS cuffed endotracheal tubes, reference RIM-70, due to the possibility of airway injuries from a sharp, rigid fragment of material in the weld line area at the distal end of the tube.

Official notice

10 Aug 2026

Last checked

26 Oct 2026

RecallScope risk estimate:Not specifiedAEMPS

An official AEMPS notice concerns Recall of lots 262401008E and 262508007E of ZARYS cuffed endotracheal tubes, reference RIM-70, due to the possibility of airway injuries from a sharp, rigid fragment of material in the weld line area at the distal end of the tube.. Recall of lots 262401008E and 262508007E of ZARYS cuffed endotracheal tubes, reference RIM-70, due to the possibility of airway injuries from a sharp, rigid fragment of material in the weld line area at the distal end of the tube. Published 2026-08-10.

What to do now

Consumer instructions are not currently available in RecallScope. Check the official notice for what to do.

Recall number

2026-388

Official notice

2026-08-10

The hazard

Recall of lots 262401008E and 262508007E of ZARYS cuffed endotracheal tubes, reference RIM-70, due to the possibility of airway injuries from a sharp, rigid fragment of material in the weld line area at the distal end of the tube.

How to check if you're affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Other

PS/AVG/141777

Raw Text

262401008E, 262508007E

Part Numbers

RIM-70 de ZARYS, RIM-70

Recall Numbers

2026-388

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Tubo endotraqueal con manguito. Número de referencia: RIM-70.

Hazard

Recall of lots 262401008E and 262508007E of ZARYS cuffed endotracheal tubes, reference RIM-70, due to the possibility of airway injuries from a sharp, rigid fragment of material in the weld line area at the distal end of the tube.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

ES · Spain

Affected area 1

ES · Spain

Source & status

Source

Tier

official

Category

HEALTH

Official classification

AEMPS — Alerta de vigilancia de productos sanitarios (tipo 2A)

Recall details

Recall number

2026-388

Event id

2026-388

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 10 Aug 2026. The next one won't wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button