Recall of certain models and serial numbers of Platinium DF4 model ICDs and CRT-Ds due to the possibility of intermittent loss of contact in the DF4 connectors, which could result in the patient not receiving the necessary therapy.

Official notice

23 Mar 2018

Last checked

26 Oct 2026

RecallScope risk estimate:Not specifiedAEMPS

An official AEMPS notice concerns Recall of certain models and serial numbers of Platinium DF4 model ICDs and CRT-Ds due to the possibility of intermittent loss of contact in the DF4 connectors, which could result in the patient not receiving the necessary therapy.. Recall of certain models and serial numbers of Platinium DF4 model ICDs and CRT-Ds due to the possibility of intermittent loss of contact in the DF4 connectors, which could result in the patient not receiving the necessary therapy. Follow up patients with implanted devices in accordance with the instructions for use, one month after discharge, and then every three months until the replacement date approaches. RECOMMENDATIONS After consulting the Spanish Society of Cardiology (SEC), AEMPS considers the company's general recommendations for following up patients with implanted devices appropriate, emphasizing the importance of including patients in the remote monitoring program through the SMARTVIEW system. Published 2018-03-23.

What to do now

After consulting the Spanish Society of Cardiology (SEC), AEMPS considers the company's general recommendations for following up patients with implanted devices appropriate, emphasizing the importance of including patients in the remote monitoring program through the SMARTVIEW system.

Recall number

2018-154

Official notice

2018-03-23

The hazard

Recall of certain models and serial numbers of Platinium DF4 model ICDs and CRT-Ds due to the possibility of intermittent loss of contact in the DF4 connectors, which could result in the patient not receiving the necessary therapy. Follow up patients with implanted devices in accordance with the instructions for use, one month after discharge, and then every three months until the replacement date approaches. RECOMMENDATIONS After consulting the Spanish Society of Cardiology (SEC), AEMPS considers the company's general recommendations for following up patients with implanted devices appropriate, emphasizing the importance of including patients in the remote monitoring program through the SMARTVIEW system.

How to check if you're affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Other

PS/FFG/44494

Models

CRT-DS PLATINIUM MODELOS DF4, DF4: VR 1240, DR 1540, CRT-D 1741, SONR CRT-D1841, 4LV CRT-D 1744, 4LV SONR CRT-D 1844, 4LV CRT-

Raw Text

La empresa indicara a cada centro en su nota de aviso los nos de serie de los dispositivos afectados.

Recall Numbers

2018-154

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Desfibriladores implantables cardiacos (DAIs) y desfibriladores con terapia de resincronización cardiaca (CRT-Ds) Platinium, modelos DF4: VR 1240, DR 1540, CRT-D 1741, SonR CRT-D1841, 4LV CRT-D 1744, 4LV SonR CRT-D 1844.

Hazard

Recall of certain models and serial numbers of Platinium DF4 model ICDs and CRT-Ds due to the possibility of intermittent loss of contact in the DF4 connectors, which could result in the patient not receiving the necessary therapy. Follow up patients with implanted devices in accordance with the instructions for use, one month after discharge, and then every three months until the replacement date approaches. RECOMMENDATIONS After consulting the Spanish Society of Cardiology (SEC), AEMPS considers the company's general recommendations for following up patients with implanted devices appropriate, emphasizing the importance of including patients in the remote monitoring program through the SMARTVIEW system.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

ES · Spain

Affected area 1

ES · Spain

Source & status

Source

Tier

official

Category

HEALTH

Official classification

AEMPS — Alerta de vigilancia de productos sanitarios (tipo 1A)

Recall details

Recall number

2018-154

Event id

2018-154

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 23 Mar 2018. The next one won't wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button