Open official notice

Recall of certain lots of certain sizes of VITUS-Fi distal fibula nails, due to possible deviations in the proximal curvature of the 4th nail caused by a manufacturing error.

Official notice

27 Oct 2025

Last checked

26 Oct 2026

RecallScope risk estimate:Not specifiedAEMPS

An official AEMPS notice concerns Recall of certain lots of certain sizes of VITUS-Fi distal fibula nails, due to possible deviations in the proximal curvature of the 4th nail caused by a manufacturing error.. Recall of certain lots of certain sizes of VITUS-Fi distal fibula nails, due to possible deviations in the proximal curvature of the 4th nail caused by a manufacturing error. Published 2025-10-27.

What to do now

After consulting its network of experts in orthopedic surgery and traumatology, AEMPS considers the recall actions taken by the company appropriate and recommends a specific (radiological) evaluation at 6 months to confirm that the fracture has healed without complications.

Recall number

2025-547

Official notice

2025-10-27

The hazard

Recall of certain lots of certain sizes of VITUS-Fi distal fibula nails, due to possible deviations in the proximal curvature of the 4th nail caused by a manufacturing error.

How to check if you're affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Other

PS/PRV/127254

Recall Numbers

2025-547

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Clavos VITUS-Fi para peroné distal.

Hazard

Recall of certain lots of certain sizes of VITUS-Fi distal fibula nails, due to possible deviations in the proximal curvature of the 4th nail caused by a manufacturing error.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

ES · Spain

Affected area 1

ES · Spain

Source & status

Source

Tier

official

Category

HEALTH

Official classification

AEMPS — Alerta de vigilancia de productos sanitarios (tipo 1A)

Recall details

Recall number

2025-547

Event id

2025-547

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 27 Oct 2025. The next one won't wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button