Open official notice

Recall of certain distal lateral fibular locking plates because threads may be missing from the plate holes, preventing the locking screws from fitting correctly.

Official notice

29 Apr 2024

Last checked

26 Oct 2026

RecallScope risk estimate:Not specifiedAEMPS

An official AEMPS notice concerns Recall of certain distal lateral fibular locking plates because threads may be missing from the plate holes, preventing the locking screws from fitting correctly.. Recall of certain distal lateral fibular locking plates because threads may be missing from the plate holes, preventing the locking screws from fitting correctly. Published 2024-04-29.

What to do now

AEMPS, having consulted its network of experts in orthopedic surgery and traumatology, considers the actions taken by the company and the recommendations for follow-up of patients with implants to be appropriate.

Recall number

2024-219

Official notice

2024-04-29

The hazard

Recall of certain distal lateral fibular locking plates because threads may be missing from the plate holes, preventing the locking screws from fitting correctly.

How to check if you're affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Other

PS/MJF/104873

Part Numbers

00-2357-000-14, 00-2357-

Recall Numbers

2024-219

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Placas de bloqueo para peroné distal lateral.

Product 2

Las placas pueden estar incluidas en set Distal Locking Lateral Fibular Plate con referencia 00-2357-000-14.

Hazard

Recall of certain distal lateral fibular locking plates because threads may be missing from the plate holes, preventing the locking screws from fitting correctly.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

ES · Spain

Affected area 1

ES · Spain

Source & status

Source

Tier

official

Category

HEALTH

Official classification

AEMPS — Alerta de vigilancia de productos sanitarios (tipo 1A)

Recall details

Recall number

2024-219

Event id

2024-219

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 29 Apr 2024. The next one won't wait for you to search.