Open official notice

Cook Medical CXI Support Catheter 2.6FR curved catheter and 4.0FR straight catheter, for cardiovascular use

Official notice

18 Nov 2015

RecallScope risk estimate:MediumFDA

An official FDA notice concerns Cook Medical CXI Support Catheter 2.6FR curved catheter and 4.0FR straight catheter, for cardiovascular use. Cook Medical has received six product complaints associated with reports of the 2.6FR curved catheter being mixed with the 4.0FR straight catheter and vice versa. Investigation revealed the products were mixed during processing. Published 2015-11-18.

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

38

Recall number

Z-0244-2016

Official notice

2015-11-18

The hazard

Cook Medical has received six product complaints associated with reports of the 2.6FR curved catheter being mixed with the 4.0FR straight catheter and vice versa. Investigation revealed the products were mixed during processing.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

72518

Recall Numbers

Z-0244-2016

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Cook Medical CXI Support Catheter 2.6FR curved catheter and 4.0FR straight catheter, for cardiovascular use.

Units affected

38

Hazard

Cook Medical has received six product complaints associated with reports of the 2.6FR curved catheter being mixed with the 4.0FR straight catheter and vice versa. Investigation revealed the products were mixed during processing.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States · CA

Location 2

US · United States · FL

Location 3

US · United States · ID

Location 4

US · United States · IL

Location 5

US · United States · MA

Location 6

US · United States · NY

Location 7

US · United States · OK

Location 8

US · United States · TX

Location 9

US · United States · UT

Location 10

US · United States · VA

Affected area 1

US · United States

Affected area 2

AT · Austria

Affected area 3

CA · Canada

Affected area 4

DE · Germany

Affected area 5

ES · Spain

Affected area 6

FR · France

Affected area 7

GB · United Kingdom

Affected area 8

HU · Hungary

Affected area 9

SE · Sweden

Company & contacts

Brand

Manufacturer

Cook Inc.

Phone

Not provided

Brand: Cook. Contact: Cook Inc..

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-0244-2016

Event ID

72518

Units affected

38

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 18 Nov 2015. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button