Open official notice

Stop using lot 1214UE of level 3 calibration serum because the alkaline phosphatase (ALP) target value for optimized AMP according to the IFCC 37°C method has been incorrectly assigned for RX series instruments by approximately 10 %.

Official notice

12 Dec 2023

Last checked

26 Oct 2026

RecallScope risk estimate:Not specifiedAEMPS

An official AEMPS notice concerns Stop using lot 1214UE of level 3 calibration serum because the alkaline phosphatase (ALP) target value for optimized AMP according to the IFCC 37°C method has been incorrectly assigned for RX series instruments by approximately 10 %.. Stop using lot 1214UE of level 3 calibration serum because the alkaline phosphatase (ALP) target value for optimized AMP according to the IFCC 37°C method has been incorrectly assigned for RX series instruments by approximately 10 %. Published 2023-12-12.

What to do now

Consumer instructions are not currently available in RecallScope. Check the official notice for what to do.

Recall number

2023-573

Official notice

2023-12-12

The hazard

Stop using lot 1214UE of level 3 calibration serum because the alkaline phosphatase (ALP) target value for optimized AMP according to the IFCC 37°C method has been incorrectly assigned for RX series instruments by approximately 10 %.

How to check if you're affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Other

PS/PL/100150

Raw Text

1214UE.

Recall Numbers

2023-573

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Suero de Calibración nivel 3, número de catálogo CAL2351.

Hazard

Stop using lot 1214UE of level 3 calibration serum because the alkaline phosphatase (ALP) target value for optimized AMP according to the IFCC 37°C method has been incorrectly assigned for RX series instruments by approximately 10 %.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

ES · Spain

Affected area 1

ES · Spain

Source & status

Source

Tier

official

Category

HEALTH

Official classification

AEMPS — Alerta de vigilancia de productos sanitarios (tipo 2A)

Recall details

Recall number

2023-573

Event id

2023-573

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 12 Dec 2023. The next one won't wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.