Open official notice

Stop using and recall the LATITUDE® programmer model 3300 because the pacing system analyzer (PSA) may deliver unexpected cross-chamber pacing in the right ventricle, potentially triggering a tachyarrhythmia.

Official notice

20 Sept 2018

Last checked

26 Oct 2026

RecallScope risk estimate:Not specifiedAEMPS

An official AEMPS notice concerns Stop using and recall the LATITUDE® programmer model 3300 because the pacing system analyzer (PSA) may deliver unexpected cross-chamber pacing in the right ventricle, potentially triggering a tachyarrhythmia.. Stop using and recall the LATITUDE® programmer model 3300 because the pacing system analyzer (PSA) may deliver unexpected cross-chamber pacing in the right ventricle, potentially triggering a tachyarrhythmia. Published 2018-09-20.

What to do now

AEMPS, having consulted the Spanish Society of Cardiology (SEC), considers the company's recommendations appropriate.

Recall number

2018-490

Official notice

2018-09-20

The hazard

Stop using and recall the LATITUDE® programmer model 3300 because the pacing system analyzer (PSA) may deliver unexpected cross-chamber pacing in the right ventricle, potentially triggering a tachyarrhythmia.

How to check if you're affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Other

PS/FFG/47787

Models

3300

Recall Numbers

2018-490

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Programador LATITUDE®, Modelo 3300

Hazard

Stop using and recall the LATITUDE® programmer model 3300 because the pacing system analyzer (PSA) may deliver unexpected cross-chamber pacing in the right ventricle, potentially triggering a tachyarrhythmia.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

ES · Spain

Affected area 1

ES · Spain

Source & status

Source

Tier

official

Category

HEALTH

Official classification

AEMPS — Alerta de productos sanitarios (tipo 1A)

Recall details

Recall number

2018-490

Event id

2018-490

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 20 Sept 2018. The next one won't wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.