Cessation of use and market withdrawal of several references and lots of the 2M 15° insert for the LUMiC® and RS cup of the MUTARS® acetabular system due to the possibility that the reinforcement ring was not correctly assembled, and recommendations for follow-up of patients implanted with this product.

Official notice

18 Jan 2018

Last checked

26 Oct 2026

RecallScope risk estimate:Not specifiedAEMPS

An official AEMPS notice concerns Cessation of use and market withdrawal of several references and lots of the 2M 15° insert for the LUMiC® and RS cup of the MUTARS® acetabular system due to the possibility that the reinforcement ring was not correctly assembled, and recommendations for follow-up of patients implanted with this product.. Cessation of use and market withdrawal of several references and lots of the 2M 15° insert for the LUMiC® and RS cup of the MUTARS® acetabular system due to the possibility that the reinforcement ring was not correctly assembled, and recommendations for follow-up of patients implanted with this product. Published 2018-01-18.

What to do now

Having consulted its network of experts in orthopedic surgery and traumatology, AEMPS considers the company's general measures appropriate. However, regarding follow-up of patients implanted with the affected product, it recommends follow-up every 3 months during the first year after implantation to detect possible complications. If no complications are detected in the first year, an annual check-up is recommended.

Recall number

2018-025

Official notice

2018-01-18

The hazard

Cessation of use and market withdrawal of several references and lots of the 2M 15° insert for the LUMiC® and RS cup of the MUTARS® acetabular system due to the possibility that the reinforcement ring was not correctly assembled, and recommendations for follow-up of patients implanted with this product.

How to check if you're affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Other

PS/SHC/42704

Recall Numbers

2018-025

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Inserto 2M 15° para copa LUMiC® y RS del sistema de revisión MUTARS® (Modular Universal Tumour And Revision System) con revestimiento de nitruro de titanio.

Hazard

Cessation of use and market withdrawal of several references and lots of the 2M 15° insert for the LUMiC® and RS cup of the MUTARS® acetabular system due to the possibility that the reinforcement ring was not correctly assembled, and recommendations for follow-up of patients implanted with this product.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

ES · Spain

Affected area 1

ES · Spain

Source & status

Source

Tier

official

Category

HEALTH

Official classification

AEMPS — Alerta de productos sanitarios (tipo 1A)

Recall details

Recall number

2018-025

Event id

2018-025

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 18 Jan 2018. The next one won't wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button