Open official notice

Stop using and recall certain references and lots of the collagen-based meniscal implant because they were shipped without the required temperature-controlled packaging and may therefore have been exposed to high temperatures during transport, in which case the implant may not provide sufficient soft tissue reinforcement.

Official notice

7 Mar 2018

Last checked

26 Oct 2026

RecallScope risk estimate:Not specifiedAEMPS

An official AEMPS notice concerns Stop using and recall certain references and lots of the collagen-based meniscal implant because they were shipped without the required temperature-controlled packaging and may therefore have been exposed to high temperatures during transport, in which case the implant may not provide sufficient soft tissue reinforcement.. Stop using and recall certain references and lots of the collagen-based meniscal implant because they were shipped without the required temperature-controlled packaging and may therefore have been exposed to high temperatures during transport, in which case the implant may not provide sufficient soft tissue reinforcement. Published 2018-03-07.

What to do now

Having consulted its network of experts in orthopedic surgery and traumatology, the AEMPS considers the company's general measures appropriate. However, regarding follow-up of patients implanted with the affected product, it recommends clinical follow-up at least at 3, 6 and 12 months after implantation, and an MRI at 6 or 12 months, which could be limited to patients presenting with effusion, inflammation or locking.

Recall number

2018-120

Official notice

2018-03-07

The hazard

Stop using and recall certain references and lots of the collagen-based meniscal implant because they were shipped without the required temperature-controlled packaging and may therefore have been exposed to high temperatures during transport, in which case the implant may not provide sufficient soft tissue reinforcement.

How to check if you're affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Other

PS/ME/43752

Part Numbers

4601, 4607, 4612

Recall Numbers

2018-120

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

CMI - Implante meniscal de base de colágeno, referencias 4601, 4607 y 4612.

Hazard

Stop using and recall certain references and lots of the collagen-based meniscal implant because they were shipped without the required temperature-controlled packaging and may therefore have been exposed to high temperatures during transport, in which case the implant may not provide sufficient soft tissue reinforcement.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

ES · Spain

Affected area 1

ES · Spain

Source & status

Source

Tier

official

Category

HEALTH

Official classification

AEMPS — Alerta de vigilancia de productos sanitarios (tipo 1A)

Recall details

Recall number

2018-120

Event id

2018-120

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 7 Mar 2018. The next one won't wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button