Open official notice

Stop using lot 517451 of the ammonia reagent kit, reference 04802290, used

Official notice

17 Nov 2020

Last checked

26 Oct 2026

RecallScope risk estimate:MediumAEMPS

An official AEMPS notice concerns Stop using lot 517451 of the ammonia reagent kit, reference 04802290, used. Stop using lot 517451 of the ammonia reagent kit, reference 04802290, used in ADVIA® Chemistry 1800, ADVIA Chemistry 2400 and ADVIA Chemistry XPT systems, because it shows a positive bias in patient plasma samples as it was manufactured with a lot of glutamate dehydrogenase (GLDH) enzyme contaminated with urease. Published 2020-11-17.

What to do now

Consumer instructions are not currently available in RecallScope. Check the official notice for what to do.

Recall number

2020-641

Official notice

2020-11-17

The hazard

Stop using lot 517451 of the ammonia reagent kit, reference 04802290, used in ADVIA® Chemistry 1800, ADVIA Chemistry 2400 and ADVIA Chemistry XPT systems, because it shows a positive bias in patient plasma samples as it was manufactured with a lot of glutamate dehydrogenase (GLDH) enzyme contaminated with urease.

How to check if you're affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Other

PS/MJF/63517

Raw Text

517451

Part Numbers

04802290

Recall Numbers

2020-641

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Kit de reactivo de amoniaco (AMM) de los Sistemas ADVIA® Chemistry, número de referencia 04802290, número de material de Siemens (SMN) 10286035

Hazard

Stop using lot 517451 of the ammonia reagent kit, reference 04802290, used in ADVIA® Chemistry 1800, ADVIA Chemistry 2400 and ADVIA Chemistry XPT systems, because it shows a positive bias in patient plasma samples as it was manufactured with a lot of glutamate dehydrogenase (GLDH) enzyme contaminated with urease.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

ES · Spain

Affected area 1

ES · Spain

Source & status

Source

Tier

official

Category

HEALTH

Official classification

AEMPS — Alerta de vigilancia de productos sanitarios (tipo 2A)

Recall details

Recall number

2020-641

Event id

2020-641

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 17 Nov 2020. The next one won't wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.