Open official notice

Safety warnings concerning EMBLEM™ subcutaneous implantable cardioverter-defibrillators (S-ICD) (Models A209 and A219) and the EMBLEM S-ICD subcutaneous electrode model 3501.

Official notice

11 Dec 2020

Last checked

26 Oct 2026

RecallScope risk estimate:MediumAEMPS

An official AEMPS notice concerns Safety warnings concerning EMBLEM™ subcutaneous implantable cardioverter-defibrillators (S-ICD) (Models A209 and A219) and the EMBLEM S-ICD subcutaneous electrode model 3501.. Safety warnings concerning EMBLEM™ subcutaneous implantable cardioverter-defibrillators (S-ICD) (Models A209 and A219) and the EMBLEM S-ICD subcutaneous electrode model 3501. Published 2020-12-11.

What to do now

After consulting the Spanish Society of Cardiology (SEC), AEMPS considers the company's recommendations for follow-up of patients implanted with the affected EMBLEM (S-ICD) devices to be appropriate. The decision for each patient must be individualized and must always be made by the physician responsible. Regarding the EMBLEM S-ICD subcutaneous electrodes: During follow-up of patients who already have implants, particular emphasis is recommended on temporarily programming all sensing vectors and assessing the quality of the detected signal and the presence of noise, both at rest and with movement of the electrode and the patient. For candidates for a first S-ICD implantation, information about the possibility of electrode failure and uncertainty about its long-term performance must be taken into account; consequently, individual assessment of the indication is recommended.

Recall number

2020-683

Official notice

2020-12-11

The hazard

Safety warnings concerning EMBLEM™ subcutaneous implantable cardioverter-defibrillators (S-ICD) (Models A209 and A219) and the EMBLEM S-ICD subcutaneous electrode model 3501.

How to check if you're affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Other

PS/CGG/54201- PS/CGG/64016- PS/CGG/64017

Models

A209, A219), A209 Y, FABRICADOS ENTRE MAYO DE 2015, DICIEMBRE DE 2017

Recall Numbers

2020-683

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Desfibriladores automáticos implantables subcutáneos (S-ICD) EMBLEM™ (Modelos A209 y A219) y Electrodo subcutáneo S-ICD EMBLEM modelo 3501

Hazard

Safety warnings concerning EMBLEM™ subcutaneous implantable cardioverter-defibrillators (S-ICD) (Models A209 and A219) and the EMBLEM S-ICD subcutaneous electrode model 3501.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

ES · Spain

Affected area 1

ES · Spain

Source & status

Source

Tier

official

Category

HEALTH

Official classification

AEMPS — Alerta de vigilancia de productos sanitarios (tipo 1A)

Recall details

Recall number

2020-683

Event id

2020-683

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 11 Dec 2020. The next one won't wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button