Open official notice

Safety warnings concerning certain Alinity hq analyzers, reference 09P68-01, because if the BD Microtainer MAP pediatric sample tube with 1.0 mg dipotassium EDTA is processed in closed mode, the system invalidates the results, and because, once the instrument has processed approximately 5,000 samples, it begins displaying zeros on the results screen and the alerts appearing in the additional information tab of the results details screen in the system control center (CCS) are invalidated.

Official notice

22 Jan 2020

Last checked

26 Oct 2026

RecallScope risk estimate:Not specifiedAEMPS

An official AEMPS notice concerns Safety warnings concerning certain Alinity hq analyzers, reference 09P68-01, because if the BD Microtainer MAP pediatric sample tube with 1.0 mg dipotassium EDTA is processed in closed mode, the system invalidates the results, and because, once the instrument has processed approximately 5,000 samples, it begins displaying zeros on the results screen and the alerts appearing in the additional information tab of the results details screen in the system control center (CCS) are invalidated.. Safety warnings concerning certain Alinity hq analyzers, reference 09P68-01, because if the BD Microtainer MAP pediatric sample tube with 1.0 mg dipotassium EDTA is processed in closed mode, the system invalidates the results, and because, once the instrument has processed approximately 5,000 samples, it begins displaying zeros on the results screen and the alerts appearing in the additional information tab of the results details screen in the system control center (CCS) are invalidated. Published 2020-01-22.

What to do now

Consumer instructions are not currently available in RecallScope. Check the official notice for what to do.

Recall number

2020-029

Official notice

2020-01-22

The hazard

Safety warnings concerning certain Alinity hq analyzers, reference 09P68-01, because if the BD Microtainer MAP pediatric sample tube with 1.0 mg dipotassium EDTA is processed in closed mode, the system invalidates the results, and because, once the instrument has processed approximately 5,000 samples, it begins displaying zeros on the results screen and the alerts appearing in the additional information tab of the results details screen in the system control center (CCS) are invalidated.

How to check if you're affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Other

PS/PF/56747

Part Numbers

09P68-01

Recall Numbers

2020-029

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Analizador Alinity hq, referencia 09P68-01

Hazard

Safety warnings concerning certain Alinity hq analyzers, reference 09P68-01, because if the BD Microtainer MAP pediatric sample tube with 1.0 mg dipotassium EDTA is processed in closed mode, the system invalidates the results, and because, once the instrument has processed approximately 5,000 samples, it begins displaying zeros on the results screen and the alerts appearing in the additional information tab of the results details screen in the system control center (CCS) are invalidated.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

ES · Spain

Affected area 1

ES · Spain

Source & status

Source

Tier

official

Category

HEALTH

Official classification

AEMPS — Alerta de vigilancia de productos sanitarios (tipo 2A)

Recall details

Recall number

2020-029

Event id

2020-029

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 22 Jan 2020. The next one won't wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.