Safety warnings concerning VG70 ventilators to inform that

Official notice

23 Sept 2020

Last checked

26 Oct 2026

RecallScope risk estimate:Not specifiedAEMPS

An official AEMPS notice concerns Safety warnings concerning VG70 ventilators to inform that. Safety warnings concerning VG70 ventilators to inform that: When the equipment starts up, the message “Data Lost 3008” may appear; this is not an alarm but an event log entry that does not affect the equipment's performance or safety. Processing information has not been provided in detail in the instructions for use, particularly regarding cleaning and disinfection of the expiratory valve, so the instructions for use have been updated to comply with the requirements of ISO 17664 “Processing of medical devices. Information to be provided by the medical device manufacturer for the processing of medical devices”. CO2 measurement is not activated because the CO2 module is an optional accessory. The CO2 module must be purchased to activate it. Published 2020-09-23.

What to do now

Consumer instructions are not currently available in RecallScope. Check the official notice for what to do.

Recall number

2020-555

Official notice

2020-09-23

The hazard

Safety warnings concerning VG70 ventilators to inform that: When the equipment starts up, the message “Data Lost 3008” may appear; this is not an alarm but an event log entry that does not affect the equipment's performance or safety. Processing information has not been provided in detail in the instructions for use, particularly regarding cleaning and disinfection of the expiratory valve, so the instructions for use have been updated to comply with the requirements of ISO 17664 “Processing of medical devices. Information to be provided by the medical device manufacturer for the processing of medical devices”. CO2 measurement is not activated because the CO2 module is an optional accessory. The CO2 module must be purchased to activate it.

How to check if you're affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Other

PS/PF/61872

Recall Numbers

2020-555

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Ventilador VG70

Hazard

Safety warnings concerning VG70 ventilators to inform that: When the equipment starts up, the message “Data Lost 3008” may appear; this is not an alarm but an event log entry that does not affect the equipment's performance or safety. Processing information has not been provided in detail in the instructions for use, particularly regarding cleaning and disinfection of the expiratory valve, so the instructions for use have been updated to comply with the requirements of ISO 17664 “Processing of medical devices. Information to be provided by the medical device manufacturer for the processing of medical devices”. CO2 measurement is not activated because the CO2 module is an optional accessory. The CO2 module must be purchased to activate it.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

ES · Spain

Affected area 1

ES · Spain

Source & status

Source

Tier

official

Category

HEALTH

Official classification

AEMPS — Alerta de vigilancia de productos sanitarios (tipo 2B)

Recall details

Recall number

2020-555

Event id

2020-555

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 23 Sept 2020. The next one won't wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button