Image not available in RecallScope.
Safety warning and update to the Model A810 pump software application
Official notice
20 Jul 2020
Last checked
26 Oct 2026
An official AEMPS notice concerns Safety warning and update to the Model A810 pump software application. Safety warning and update to version 1.1.342 of the Model A810 software application for the SynchroMed II pump, used with the CT900 clinician programming tablet, because earlier software versions do not resolve a specific pump memory error that could potentially prevent execution of the pump update command. Published 2020-07-20.
What to do now
Consumer instructions are not currently available in RecallScope. Check the official notice for what to do.
Recall number
2020-405
Official notice
2020-07-20
The hazard
The hazard
Safety warning and update to version 1.1.342 of the Model A810 software application for the SynchroMed II pump, used with the CT900 clinician programming tablet, because earlier software versions do not resolve a specific pump memory error that could potentially prevent execution of the pump update command.
How to check if you're affected
How to check if you're affected
Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.
Other
Models
Recall Numbers
Checked manually this time?
Follow brands and keywords in the app to receive alerts for matching official notices.
Turn on recall alertsAffected products
Affected products
Product 1
Hazard
Where it was sold
Where it was sold
Retailer information is not currently available in RecallScope. Check the official notice for distribution details.
Location 1
Affected area 1
Source & status
Source & status
Source
Tier
Category
Official classification
Recall details
Recall details
Recall number
Event id
Source tier
Documents, notices & source
See the official notice and source documents for full details.