Open official notice

Safety warning concerning fluid protein assays

Official notice

11 Oct 2021

Last checked

26 Oct 2026

RecallScope risk estimate:Not specifiedAEMPS

An official AEMPS notice concerns Safety warning concerning fluid protein assays. Safety warning concerning urinary/cerebrospinal fluid protein (UCFP) assays for Atellica CH 930 and ADVIA Chemistry 1800/2400 and XPT, due to the possibility of carryover of these reagents that could affect enzymatic hemoglobin A1c (A1c_E and A1c_EM) results and lead to falsely decreased results when this assay is run after the UCFP assay. Published 2021-10-11.

What to do now

Consumer instructions are not currently available in RecallScope. Check the official notice for what to do.

Recall number

2021-560

Official notice

2021-10-11

The hazard

Safety warning concerning urinary/cerebrospinal fluid protein (UCFP) assays for Atellica CH 930 and ADVIA Chemistry 1800/2400 and XPT, due to the possibility of carryover of these reagents that could affect enzymatic hemoglobin A1c (A1c_E and A1c_EM) results and lead to falsely decreased results when this assay is run after the UCFP assay.

How to check if you're affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Other

PS/JMP/72952

Recall Numbers

2021-560

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Ensayo de proteína del líquido urinario/cefalorraquídeo (UCFP): Para Atellica CH 930 (nº de material de Siemens: 11097543); Para ADVIA Chemistry 1800/2400 y XPT (nº de material de Siemens: 11319151)

Hazard

Safety warning concerning urinary/cerebrospinal fluid protein (UCFP) assays for Atellica CH 930 and ADVIA Chemistry 1800/2400 and XPT, due to the possibility of carryover of these reagents that could affect enzymatic hemoglobin A1c (A1c_E and A1c_EM) results and lead to falsely decreased results when this assay is run after the UCFP assay.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

ES · Spain

Affected area 1

ES · Spain

Source & status

Source

Tier

official

Category

HEALTH

Official classification

AEMPS — Alerta de vigilancia de productos sanitarios (tipo 2A)

Recall details

Recall number

2021-560

Event id

2021-560

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 11 Oct 2021. The next one won't wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button