Open official notice

Safety warning concerning ORTHO VISION and ORTHO VISION Max analyzers configured with software versions 5.12.8 or 5.13.0, due to the possibility that, upon restarting, an unexpected discharge of saline solution into a reagent/diluent vial may occur after the pipette arm (PIPA) exits idle mode (IDLE), which could lead to contamination of the sample, dilution tray or reagent and consequently produce false-negative or false-positive test results.

Official notice

29 Dec 2020

Last checked

26 Oct 2026

RecallScope risk estimate:HighAEMPS

An official AEMPS notice concerns Safety warning concerning ORTHO VISION and ORTHO VISION Max analyzers configured with software versions 5.12.8 or 5.13.0, due to the possibility that, upon restarting, an unexpected discharge of saline solution into a reagent/diluent vial may occur after the pipette arm (PIPA) exits idle mode (IDLE), which could lead to contamination of the sample, dilution tray or reagent and consequently produce false-negative or false-positive test results.. Safety warning concerning ORTHO VISION and ORTHO VISION Max analyzers configured with software versions 5.12.8 or 5.13.0, due to the possibility that, upon restarting, an unexpected discharge of saline solution into a reagent/diluent vial may occur after the pipette arm (PIPA) exits idle mode (IDLE), which could lead to contamination of the sample, dilution tray or reagent and consequently produce false-negative or false-positive test results. Published 2020-12-29.

What to do now

Consumer instructions are not currently available in RecallScope. Check the official notice for what to do.

Recall number

2020-728

Official notice

2020-12-29

The hazard

Safety warning concerning ORTHO VISION and ORTHO VISION Max analyzers configured with software versions 5.12.8 or 5.13.0, due to the possibility that, upon restarting, an unexpected discharge of saline solution into a reagent/diluent vial may occur after the pipette arm (PIPA) exits idle mode (IDLE), which could lead to contamination of the sample, dilution tray or reagent and consequently produce false-negative or false-positive test results.

How to check if you're affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Other

PS/PPR/64590

Recall Numbers

2020-728

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Analizador ORTHO VISION®, Código de producto 6904579 y Analizador ORTHO VISION® Max, Código de producto 6904578, para cassettes ORTHO BioVue®

Hazard

Safety warning concerning ORTHO VISION and ORTHO VISION Max analyzers configured with software versions 5.12.8 or 5.13.0, due to the possibility that, upon restarting, an unexpected discharge of saline solution into a reagent/diluent vial may occur after the pipette arm (PIPA) exits idle mode (IDLE), which could lead to contamination of the sample, dilution tray or reagent and consequently produce false-negative or false-positive test results.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

ES · Spain

Affected area 1

ES · Spain

Source & status

Source

Tier

official

Category

HEALTH

Official classification

AEMPS — Alerta de vigilancia de productos sanitarios (tipo 2A)

Recall details

Recall number

2020-728

Event id

2020-728

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 29 Dec 2020. The next one won't wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button