Open official notice

Safety warning concerning the possibility that external defibrillators

Official notice

10 Jul 2020

Last checked

26 Oct 2026

RecallScope risk estimate:MediumAEMPS

An official AEMPS notice concerns Safety warning concerning the possibility that external defibrillators. Safety warning concerning the possibility that LIFEPAK 500 automated external defibrillators may fail to detect a patient connection due to wear of their components. Published 2020-07-10.

What to do now

AEMPS recommends that each facility assess the benefit/risk balance associated with using this product and consider replacing it.

Recall number

2020-382

Official notice

2020-07-10

The hazard

Safety warning concerning the possibility that LIFEPAK 500 automated external defibrillators may fail to detect a patient connection due to wear of their components.

How to check if you're affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Other

PS/MTM/59814

Raw Text

Todos los números de serie de los productos cuya referencia empieza por 99401-(XXXXXX): todos los, DEA LIFEPAK 500 que permanecen en uso.

Part Numbers

empieza por 99401-(XXXXXX): todos los

Recall Numbers

2020-382

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Desfibrilador externo automático (DEA) LIFEPAK® 500

Hazard

Safety warning concerning the possibility that LIFEPAK 500 automated external defibrillators may fail to detect a patient connection due to wear of their components.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

ES · Spain

Affected area 1

ES · Spain

Source & status

Source

Tier

official

Category

HEALTH

Official classification

AEMPS — Alerta de vigilancia de productos sanitarios (tipo 1A)

Recall details

Recall number

2020-382

Event id

2020-382

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 10 Jul 2020. The next one won't wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button