Open official notice

Safety warning concerning the quality control (QC) function of Alinity ci-series systems because, following an assay file update

Official notice

2 Jul 2020

Last checked

26 Oct 2026

RecallScope risk estimate:MediumAEMPS

An official AEMPS notice concerns Safety warning concerning the quality control (QC) function of Alinity ci-series systems because, following an assay file update. Safety warning concerning the quality control (QC) function of Alinity ci-series systems because, following an assay file update: 1. Subsequently generated QC results are not evaluated for Westgard rule violations when the system is configured to evaluate these rules. 2. Sample results generated after a quality control failure are not flagged with a CNTL alert, regardless of the control configuration (QC range, mean/SD, Westgard). Published 2020-07-02.

What to do now

Consumer instructions are not currently available in RecallScope. Check the official notice for what to do.

Recall number

2020-363

Official notice

2020-07-02

The hazard

Safety warning concerning the quality control (QC) function of Alinity ci-series systems because, following an assay file update: 1. Subsequently generated QC results are not evaluated for Westgard rule violations when the system is configured to evaluate these rules. 2. Sample results generated after a quality control failure are not flagged with a CNTL alert, regardless of the control configuration (QC range, mean/SD, Westgard).

How to check if you're affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Part Numbers

03R70-01

Recall Numbers

2020-363

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Módulo de control del sistema (SCM) Alinity ci-series, referencia 03R70-01.

Hazard

Safety warning concerning the quality control (QC) function of Alinity ci-series systems because, following an assay file update: 1. Subsequently generated QC results are not evaluated for Westgard rule violations when the system is configured to evaluate these rules. 2. Sample results generated after a quality control failure are not flagged with a CNTL alert, regardless of the control configuration (QC range, mean/SD, Westgard).

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

ES · Spain

Affected area 1

ES · Spain

Source & status

Source

Tier

official

Category

HEALTH

Official classification

AEMPS — Alerta de productos sanitarios (tipo 2A)

Recall details

Recall number

2020-363

Event id

2020-363

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 2 Jul 2020. The next one won't wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button