Open official notice

Safety warning regarding the possibility of incorrect urine specific gravity results being transmitted by iChemVELOCITY Urine Chemistry Systems because the S3 alert may not be generated when the instrument cannot measure specific gravity due to interference or a hardware failure, and an update to the instructions for use because they state that the S3 alert means the value is >1.060, rather than that the result obtained is invalid.

Official notice

29 Nov 2019

Last checked

26 Oct 2026

RecallScope risk estimate:MediumAEMPS

An official AEMPS notice concerns Safety warning regarding the possibility of incorrect urine specific gravity results being transmitted by iChemVELOCITY Urine Chemistry Systems because the S3 alert may not be generated when the instrument cannot measure specific gravity due to interference or a hardware failure, and an update to the instructions for use because they state that the S3 alert means the value is >1.060, rather than that the result obtained is invalid.. Safety warning regarding the possibility of incorrect urine specific gravity results being transmitted by iChemVELOCITY Urine Chemistry Systems because the S3 alert may not be generated when the instrument cannot measure specific gravity due to interference or a hardware failure, and an update to the instructions for use because they state that the S3 alert means the value is >1.060, rather than that the result obtained is invalid. Published 2019-11-29.

What to do now

Consumer instructions are not currently available in RecallScope. Check the official notice for what to do.

Recall number

2019-633

Official notice

2019-11-29

The hazard

Safety warning regarding the possibility of incorrect urine specific gravity results being transmitted by iChemVELOCITY Urine Chemistry Systems because the S3 alert may not be generated when the instrument cannot measure specific gravity due to interference or a hardware failure, and an update to the instructions for use because they state that the S3 alert means the value is >1.060, rather than that the result obtained is invalid.

How to check if you're affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Other

PS/PF/55871

Recall Numbers

2019-633

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Sistemas iChemVELOCITY Urine Chemistry

Hazard

Safety warning regarding the possibility of incorrect urine specific gravity results being transmitted by iChemVELOCITY Urine Chemistry Systems because the S3 alert may not be generated when the instrument cannot measure specific gravity due to interference or a hardware failure, and an update to the instructions for use because they state that the S3 alert means the value is >1.060, rather than that the result obtained is invalid.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

ES · Spain

Affected area 1

ES · Spain

Source & status

Source

Tier

official

Category

HEALTH

Official classification

AEMPS — Alerta de vigilancia de productos sanitarios (tipo 2A)

Recall details

Recall number

2019-633

Event id

2019-633

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 29 Nov 2019. The next one won't wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.