Update to the safety warnings concerning HemosIL heparin calibrators, starting with lot N1157997, because value assignment has shown a bias relative to WHO international standards for unfractionated heparin (UFH) and low molecular weight heparin (LMWH), causing reduced recovery when measuring heparin in both controls and samples when used with HemosIL Liquid Anti-Xa reagent (Reference 00020302601).

Official notice

24 Mar 2026

Last checked

26 Oct 2026

RecallScope risk estimate:MediumAEMPS

An official AEMPS notice concerns Update to the safety warnings concerning HemosIL heparin calibrators, starting with lot N1157997, because value assignment has shown a bias relative to WHO international standards for unfractionated heparin (UFH) and low molecular weight heparin (LMWH), causing reduced recovery when measuring heparin in both controls and samples when used with HemosIL Liquid Anti-Xa reagent (Reference 00020302601).. Update to the safety warnings concerning HemosIL heparin calibrators, starting with lot N1157997, because value assignment has shown a bias relative to WHO international standards for unfractionated heparin (UFH) and low molecular weight heparin (LMWH), causing reduced recovery when measuring heparin in both controls and samples when used with HemosIL Liquid Anti-Xa reagent (Reference 00020302601). Published 2026-03-24.

What to do now

Consumer instructions are not currently available in RecallScope. Check the official notice for what to do.

Recall number

2026-133

Official notice

2026-03-24

The hazard

Update to the safety warnings concerning HemosIL heparin calibrators, starting with lot N1157997, because value assignment has shown a bias relative to WHO international standards for unfractionated heparin (UFH) and low molecular weight heparin (LMWH), causing reduced recovery when measuring heparin in both controls and samples when used with HemosIL Liquid Anti-Xa reagent (Reference 00020302601).

How to check if you're affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Other

UDI-DI 08426950472483, PS/RJC/115911

Raw Text

Del lote N1157997 en adelante.

Part Numbers

00020302601), 0020300600

Recall Numbers

2026-133

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

HemosIL calibradores de heparina. Referencia 0020300600. UDI-DI 08426950472483.

Hazard

Update to the safety warnings concerning HemosIL heparin calibrators, starting with lot N1157997, because value assignment has shown a bias relative to WHO international standards for unfractionated heparin (UFH) and low molecular weight heparin (LMWH), causing reduced recovery when measuring heparin in both controls and samples when used with HemosIL Liquid Anti-Xa reagent (Reference 00020302601).

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

ES · Spain

Affected area 1

ES · Spain

Source & status

Source

Tier

official

Category

HEALTH

Official classification

AEMPS — Alerta de vigilancia de productos sanitarios (tipo 2A)

Recall details

Recall number

2026-133

Event id

2026-133

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 24 Mar 2026. The next one won't wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button