Open official notice

Updated information and new corrective measures related to stopping use of FTD in vitro diagnostic kits because an internal investigation found insufficient data to support the performance claims in the Instructions for Use.

Official notice

7 Feb 2020

Last checked

26 Oct 2026

RecallScope risk estimate:Not specifiedAEMPS

An official AEMPS notice concerns Updated information and new corrective measures related to stopping use of FTD in vitro diagnostic kits because an internal investigation found insufficient data to support the performance claims in the Instructions for Use.. Updated information and new corrective measures related to stopping use of FTD in vitro diagnostic kits because an internal investigation found insufficient data to support the performance claims in the Instructions for Use. Published 2020-02-07.

What to do now

Consumer instructions are not currently available in RecallScope. Check the official notice for what to do.

Recall number

2020-078

Official notice

2020-02-07

The hazard

Updated information and new corrective measures related to stopping use of FTD in vitro diagnostic kits because an internal investigation found insufficient data to support the performance claims in the Instructions for Use.

How to check if you're affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Other

PS/PF/56710

Raw Text

Todos

Recall Numbers

2020-078

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Kits FTD. Ver listado detallado de productos y número de catálogo en las tablas 1 y 2 de la nota de aviso de la empresa.

Hazard

Updated information and new corrective measures related to stopping use of FTD in vitro diagnostic kits because an internal investigation found insufficient data to support the performance claims in the Instructions for Use.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

ES · Spain

Affected area 1

ES · Spain

Source & status

Source

Tier

official

Category

HEALTH

Official classification

AEMPS — Alerta de vigilancia de productos sanitarios (tipo 2A)

Recall details

Recall number

2020-078

Event id

2020-078

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 7 Feb 2020. The next one won't wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button