Thermocool smarttouch sf bi-directional navigation catheter, ref: d-1348-05-s. Used for catheter-based cardiac electrophysiological mapping

Official notice

9 Nov 2022

RecallScope risk estimate:MediumFDA

An official FDA notice concerns Thermocool smarttouch sf bi-directional navigation catheter, ref: d-1348-05-s. Used for catheter-based cardiac electrophysiological mapping. Manufacturing error resulted in potential contamination of Bi-Directional navigation catheter irrigation path with cellulose fibers that can cause 1) Failure of initial flushing inspection, 2) Insufficient irrigation causing catheter tip temperature increase, limiting the ability to deliver ablation, or 3) particles flushed through the catheter irrigation may have thrombogenic/toxic effect. Published 2022-11-09.

What to do now

Do not consume or use the affected product. Return it to the place of purchase or follow instructions in the official FDA notice.

Units affected

91

Recall number

Z-0169-2023

Official notice

2022-11-09

The hazard

Manufacturing error resulted in potential contamination of Bi-Directional navigation catheter irrigation path with cellulose fibers that can cause 1) Failure of initial flushing inspection, 2) Insufficient irrigation causing catheter tip temperature increase, limiting the ability to deliver ablation, or 3) particles flushed through the catheter irrigation may have thrombogenic/toxic effect.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Other

UDI-DI 10846835010183

Event Ids

90887

Recall Numbers

Z-0169-2023

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

THERMOCOOL SMARTTOUCH SF Bi-Directional Navigation Catheter, REF: D-1348-05-S. Used for catheter-based cardiac electrophysiological mapping (stimulation and recording) and, in conjunction with an RF generator, for cardiac ablation.

Units affected

91

Hazard

Manufacturing error resulted in potential contamination of Bi-Directional navigation catheter irrigation path with cellulose fibers that can cause 1) Failure of initial flushing inspection, 2) Insufficient irrigation causing catheter tip temperature increase, limiting the ability to deliver ablation, or 3) particles flushed through the catheter irrigation may have thrombogenic/toxic effect.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Affected area 2

DE · Germany

Affected area 3

HR · Croatia

Affected area 4

HU · Hungary

Company & contacts

Manufacturer

Johnson & Johnson Surgical Vision, Inc.

Phone

Not provided

Brand: Johnson Johnson Surgical Vision. Contact: Johnson & Johnson Surgical Vision, Inc..

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-0169-2023

Event ID

90887

Units affected

91

Status

active

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 9 Nov 2022. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button