Open official notice

PFC SIGMA Cruciate Retaining (CR); Cemented Femoral devices; (Size 5 RT) Used during primary total knee arthroplasty to improve patient

Official notice

5 Mar 2014

RecallScope risk estimate:HighFDA

An official FDA notice concerns PFC SIGMA Cruciate Retaining (CR); Cemented Femoral devices; (Size 5 RT) Used during primary total knee arthroplasty to improve patient. Knee implant femoral devices were found to have anomalous microstructure. Porosity and the presence of blocky carbides, within the cobalt chrome part, can decrease the overall fatigue strength. This decrease in strength could cause implants to fracture under what would normally be acceptable loads. Published 2014-03-05.

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

2

Recall number

Z-1077-2014

Official notice

2014-03-05

The hazard

Knee implant femoral devices were found to have anomalous microstructure. Porosity and the presence of blocky carbides, within the cobalt chrome part, can decrease the overall fatigue strength. This decrease in strength could cause implants to fracture under what would normally be acceptable loads.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Lots

7810268, 7806934

Event Ids

67159

Recall Numbers

Z-1077-2014

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

PFC SIGMA Cruciate Retaining (CR); Cemented Femoral devices; (Size 5/RT) Used during primary total knee arthroplasty to improve patient mobility.

Units affected

2

Hazard

Knee implant femoral devices were found to have anomalous microstructure. Porosity and the presence of blocky carbides, within the cobalt chrome part, can decrease the overall fatigue strength. This decrease in strength could cause implants to fracture under what would normally be acceptable loads.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States · CA

Location 2

US · United States · FL

Location 3

US · United States · IA

Location 4

US · United States · KY

Affected area 1

US · United States

Affected area 2

CZ · Czech Republic

Affected area 3

DE · Germany

Affected area 4

FI · Finland

Affected area 5

SE · Sweden

Company & contacts

Manufacturer

DePuy Orthopaedics, Inc.

Phone

Not provided

Brand: Depuy Orthopaedics. Contact: DePuy Orthopaedics, Inc..

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-1077-2014

Event ID

67159

Units affected

2

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 5 Mar 2014. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button