Open official notice

OptiMedica Catalys Precision Laser System, CE 0044 containing software cOS 5.00.33 Product Label for Serial Numbers: 44109015; 44106215

Official notice

10 Oct 2018

RecallScope risk estimate:MediumFDA

An official FDA notice concerns OptiMedica Catalys Precision Laser System, CE 0044 containing software cOS 5.00.33 Product Label for Serial Numbers: 44109015; 44106215. Software upgrade exhibits failures when executing Daily Alignment Verification (DAV), specifically, cuts on the plastic hemisphere during DAV do not match the intended cuts shown on the treatment overlay screen on the system GUI. Published 2018-10-10.

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

9

Recall number

Z-0066-2019

Official notice

2018-10-10

The hazard

Software upgrade exhibits failures when executing Daily Alignment Verification (DAV), specifically, cuts on the plastic hemisphere during DAV do not match the intended cuts shown on the treatment overlay screen on the system GUI.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

80849

Recall Numbers

Z-0066-2019

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

OptiMedica Catalys Precision Laser System, CE 0044 containing software cOS 5.00.33 Product Label for Serial Numbers: 44109015; 44106215; 44107315; and 44065613 label states "Manufactured for: OptiMedica Corporation 1310 Moffett Park Drive Sunnyvale, CA 94089 USA Product Label for Serial Numbers: 46120916; 46123116, 46124617; 46131418, and 46131618 label states "Manufactured for: OptiMedica Corporation 510 Cottonwood Drive Milpitas, CA 95035 USA The Catalys Precision Laser System is indicated for use in patients undergoing cataract surgery for removal of the crystalline lens. Intended uses in cataract surgery include anterior capsulotomy, phacofragmentation, and the creation of single plane and multi-plane arc cuts/incisions in the cornea, each of which may be performed either individually or consecutively during the same procedure.

Units affected

9

Hazard

Software upgrade exhibits failures when executing Daily Alignment Verification (DAV), specifically, cuts on the plastic hemisphere during DAV do not match the intended cuts shown on the treatment overlay screen on the system GUI.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States · CO

Location 2

US · United States · GA

Location 3

US · United States · IL

Location 4

US · United States · MD

Affected area 1

US · United States

Affected area 2

AT · Austria

Affected area 3

DE · Germany

Company & contacts

Manufacturer

Johnson & Johnson Surgical Vision Inc

Phone

Not provided

Brand: Johnson Johnson Surgical Vision. Contact: Johnson & Johnson Surgical Vision Inc.

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-0066-2019

Event ID

80849

Units affected

9

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 10 Oct 2018. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button