Open official notice

BrainLABs VectorVision spine is intended for use as an intraoperative image-guided localization system for minimally invasive surgery. It

Official notice

19 Feb 2014

RecallScope risk estimate:MediumFDA

An official FDA notice concerns BrainLABs VectorVision spine is intended for use as an intraoperative image-guided localization system for minimally invasive surgery. It. Brainlab has determined that pins of thee Instrument Star Unit ML (Calibration with ICM4) with specific serial numbers might have been made of an incorrect material that is not suitable for the necessary disinfection and sterilization of the product. If corrosion occurs and is not detected by the user, and the device is used during surgery, corroded particles of the pin could, directly or indirect Published 2014-02-19.

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

25

Recall number

Z-0967-2014

Official notice

2014-02-19

The hazard

Brainlab has determined that pins of thee Instrument Star Unit ML (Calibration with ICM4) with specific serial numbers might have been made of an incorrect material that is not suitable for the necessary disinfection and sterilization of the product. If corrosion occurs and is not detected by the user, and the device is used during surgery, corroded particles of the pin could, directly or indirect

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

67017

Recall Numbers

Z-0967-2014

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

BrainLAB's VectorVision spine is intended for use as an intraoperative image-guided localization system for minimally invasive surgery. It links a freehand probe, tracked by a passive marker sensor system to virtual computer image space on a patient's preoperative image data that is processed by a VectorVision workstation. The system is indicated for any medical condition in which the use of stereoscopic surgery may be appropriate and where a reference to a rigid anatomical structure, such as the skull, the pelvis, a long bone or vertebra can be identified relative to a CT, x-ray, or MR-based model of the anatomy.

Units affected

25

Hazard

Brainlab has determined that pins of thee Instrument Star Unit ML (Calibration with ICM4) with specific serial numbers might have been made of an incorrect material that is not suitable for the necessary disinfection and sterilization of the product. If corrosion occurs and is not detected by the user, and the device is used during surgery, corroded particles of the pin could, directly or indirect

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States · CA

Location 2

US · United States · DC

Location 3

US · United States · MN

Location 4

US · United States · NC

Location 5

US · United States · NY

Location 6

US · United States · OH

Location 7

US · United States · TX

Affected area 1

US · United States

Affected area 2

AE · United Arab Emirates

Affected area 3

AU · Australia

Affected area 4

CL · Chile

Affected area 5

DE · Germany

Affected area 6

HU · Hungary

Affected area 7

IT · Italy

Affected area 8

RU · Russia

Affected area 9

TH · Thailand

Company & contacts

Manufacturer

Brainlab AG

Phone

Not provided

Brand: Brainlab. Contact: Brainlab AG.

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-0967-2014

Event ID

67017

Units affected

25

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 19 Feb 2014. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button