Veritas Advanced Infusion Packs, REF: VRT-AI

Official notice

1 Feb 2023

RecallScope risk estimate:CriticalFDA

An official FDA notice concerns Veritas Advanced Infusion Packs, REF: VRT-AI. Weld protrusion/physical gap between the housing and phaco pack cover, may exceed design specification, which could lead to priming cycle failure and/or suboptimal vacuum delivered to the phacoemulsification and irrigation/aspiration handpieces during surgery, which could result in surgical delay and/or longer surgical time, which could result in transient corneal edema. Published 2023-02-01.

What to do now

Stop using the affected device and follow the corrective action in the official FDA notice. Contact your healthcare provider or the manufacturer for next steps.

Units affected

3,133

Recall number

Z-1005-2023

Official notice

2023-02-01

The hazard

Weld protrusion/physical gap between the housing and phaco pack cover, may exceed design specification, which could lead to priming cycle failure and/or suboptimal vacuum delivered to the phacoemulsification and irrigation/aspiration handpieces during surgery, which could result in surgical delay and/or longer surgical time, which could result in transient corneal edema.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Other

UDI-DI 05050474700901

Event Ids

91290

Recall Numbers

Z-1005-2023

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Veritas Advanced Infusion Packs, REF: VRT-AI

Units affected

3,133

Hazard

Weld protrusion/physical gap between the housing and phaco pack cover, may exceed design specification, which could lead to priming cycle failure and/or suboptimal vacuum delivered to the phacoemulsification and irrigation/aspiration handpieces during surgery, which could result in surgical delay and/or longer surgical time, which could result in transient corneal edema.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States · FL

Location 2

US · United States · IL

Location 3

US · United States · MN

Location 4

US · United States · MS

Location 5

US · United States · NY

Location 6

US · United States · OH

Location 7

US · United States · TN

Location 8

US · United States · TX

Affected area 1

US · United States

Affected area 2

BE · Belgium

Affected area 3

CH · Switzerland

Affected area 4

CN · China

Affected area 5

DE · Germany

Affected area 6

EE · Estonia

Affected area 7

FR · France

Affected area 8

GB · United Kingdom

Affected area 9

IT · Italy

Affected area 10

JP · Japan

Affected area 11

MY · Malaysia

Affected area 12

NL · Netherlands

Affected area 13

PT · Portugal

Affected area 14

TH · Thailand

Company & contacts

Manufacturer

Johnson & Johnson Surgical Vision, Inc.

Phone

Not provided

Brand: Johnson Johnson Surgical Vision. Contact: Johnson & Johnson Surgical Vision, Inc..

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-1005-2023

Event ID

91290

Units affected

3,133

Status

active

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 1 Feb 2023. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button