Spine & Trauma 3D Navigation 1.0 (subpart of the system Navigation Software Spine & Trauma 3D, Version 3.0 ) Intended Use The Spine &

Official notice

24 Apr 2019

RecallScope risk estimate:CriticalFDA

An official FDA notice concerns Spine & Trauma 3D Navigation 1.0 (subpart of the system Navigation Software Spine & Trauma 3D, Version 3.0 ) Intended Use The Spine &. In certain occurrences, the affected navigation software application might unexpectedly display a navigated instrument in an axial, coronal/sagittal (ACS) view representation with fixed planes in the image reconstruction (the not-updated ACS view ), instead of displaying the desirable view representation Inline View , which is commonly used for navigating invasive instruments at the spine. This could lead the surgeon to be unable to determine the position of the navigated instrument. This might occur after a crash restore or after changing between different navigation workflows during the same patient treatment. Published 2019-04-24.

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

148

Recall number

Z-1082-2019

Official notice

2019-04-24

The hazard

In certain occurrences, the affected navigation software application might unexpectedly display a navigated instrument in an axial, coronal/sagittal (ACS) view representation with fixed planes in the image reconstruction (the not-updated ACS view ), instead of displaying the desirable view representation Inline View , which is commonly used for navigating invasive instruments at the spine. This could lead the surgeon to be unable to determine the position of the navigated instrument. This might occur after a crash restore or after changing between different navigation workflows during the same patient treatment.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

82304

Recall Numbers

Z-1082-2019

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Spine & Trauma 3D Navigation 1.0 (subpart of the system Navigation Software Spine & Trauma 3D, Version 3.0 ) Intended Use The Spine & Trauma Navigation System is intended as an intraoperative image-guided localization system to enable minimally invasive surgery.

Units affected

148

Hazard

In certain occurrences, the affected navigation software application might unexpectedly display a navigated instrument in an axial, coronal/sagittal (ACS) view representation with fixed planes in the image reconstruction (the not-updated ACS view ), instead of displaying the desirable view representation Inline View , which is commonly used for navigating invasive instruments at the spine. This could lead the surgeon to be unable to determine the position of the navigated instrument. This might occur after a crash restore or after changing between different navigation workflows during the same patient treatment.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Affected area 2

AT · Austria

Affected area 3

AU · Australia

Affected area 4

BE · Belgium

Affected area 5

CA · Canada

Affected area 6

CH · Switzerland

Affected area 7

CZ · Czech Republic

Affected area 8

DE · Germany

Affected area 9

DK · Denmark

Affected area 10

ES · Spain

Affected area 11

FR · France

Affected area 12

GB · United Kingdom

Affected area 13

GR · Greece

Affected area 14

HK · Hong Kong

Affected area 15

IE · Ireland

Affected area 16

IT · Italy

Affected area 17

JP · Japan

Affected area 18

MY · Malaysia

Affected area 19

NZ · New Zealand

Affected area 20

SE · Sweden

Affected area 21

TH · Thailand

Affected area 22

TW · Taiwan

Affected area 23

ZA · South Africa

Company & contacts

Manufacturer

Brainlab AG

Phone

Not provided

Brand: Brainlab. Contact: Brainlab AG.

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class I

Recall details

Recall number

Z-1082-2019

Event ID

82304

Units affected

148

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 24 Apr 2019. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button