Open official notice

RT Elements Software revisions of the RT Elements applications have a specific software version number. Specifically the following RT

Official notice

1 May 2019

RecallScope risk estimate:MediumFDA

An official FDA notice concerns RT Elements Software revisions of the RT Elements applications have a specific software version number. Specifically the following RT. There is a potential for an incorrect dose distribution calculation by Brainlab RT Elements software (for affected versions) under specific circumstances when using the Pencil Beam algorithm on the GPU (graphics card), as is the default system setting. Published 2019-05-01.

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

164

Recall number

Z-1205-2019

Official notice

2019-05-01

The hazard

There is a potential for an incorrect dose distribution calculation by Brainlab RT Elements software (for affected versions) under specific circumstances when using the Pencil Beam algorithm on the GPU (graphics card), as is the default system setting.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

82601

Recall Numbers

Z-1205-2019

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

RT Elements Software revisions of the RT Elements applications have a specific software version number. Specifically the following RT Elements applications/versions are affected: - Cranial SRS 1.0.0 and 1.5.0 - Spine SRS 1.0.0 and 1.5.0 - Multiple Brain Mets SRS 1.5.0 - RT QA 1.0.0 and 1.5.0 Product Usage: The RT Elements are applications for radiation treatment planning for use in stereotactic, conformal, computer planned, Linac based radiation treatment of cranial, head and neck and extracranial lesions.

Units affected

164

Hazard

There is a potential for an incorrect dose distribution calculation by Brainlab RT Elements software (for affected versions) under specific circumstances when using the Pencil Beam algorithm on the GPU (graphics card), as is the default system setting.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Affected area 2

AE · United Arab Emirates

Affected area 3

AR · Argentina

Affected area 4

AT · Austria

Affected area 5

AU · Australia

Affected area 6

BE · Belgium

Affected area 7

BR · Brazil

Affected area 8

CA · Canada

Affected area 9

CH · Switzerland

Affected area 10

CN · China

Affected area 11

CO · Colombia

Affected area 12

DE · Germany

Affected area 13

ES · Spain

Affected area 14

FI · Finland

Affected area 15

FR · France

Affected area 16

GB · United Kingdom

Affected area 17

HK · Hong Kong

Affected area 18

IE · Ireland

Affected area 19

IL · Israel

Affected area 20

IT · Italy

Affected area 21

JO · Jordan

Affected area 22

JP · Japan

Affected area 23

KR · South Korea

Affected area 24

MX · Mexico

Affected area 25

NL · Netherlands

Affected area 26

PL · Poland

Affected area 27

PT · Portugal

Affected area 28

SV · El Salvador

Affected area 29

TW · Taiwan

Affected area 30

ZA · South Africa

Company & contacts

Manufacturer

Brainlab AG

Phone

Not provided

Brand: Brainlab. Contact: Brainlab AG.

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-1205-2019

Event ID

82601

Units affected

164

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 1 May 2019. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button