Reliance Vision Multi-Chamber Washer Disinfector, Reference Number FH83062

Official notice

24 May 2023

RecallScope risk estimate:HighFDA

An official FDA notice concerns Reliance Vision Multi-Chamber Washer Disinfector, Reference Number FH83062. In the remote occurrence in which the electrical contactor component present in the drying chambers of the Reliance Vision Multi-Chamber Washer/Disinfector malfunctions, the heating elements in the drying chamber could overheat, eventually resulting in smoke and/or fire. Published 2023-05-24.

What to do now

Stop using the affected device and follow the corrective action in the official FDA notice. Contact your healthcare provider or the manufacturer for next steps.

Units affected

322

Recall number

Z-1595-2023

Official notice

2023-05-24

The hazard

In the remote occurrence in which the electrical contactor component present in the drying chambers of the Reliance Vision Multi-Chamber Washer/Disinfector malfunctions, the heating elements in the drying chamber could overheat, eventually resulting in smoke and/or fire.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Other

UDI-DI 00724995193164

Event Ids

92044

Recall Numbers

Z-1595-2023

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Reliance Vision Multi-Chamber Washer/Disinfector, Reference Number FH83062

Units affected

322

Hazard

In the remote occurrence in which the electrical contactor component present in the drying chambers of the Reliance Vision Multi-Chamber Washer/Disinfector malfunctions, the heating elements in the drying chamber could overheat, eventually resulting in smoke and/or fire.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Affected area 2

AU · Australia

Affected area 3

BR · Brazil

Affected area 4

CA · Canada

Affected area 5

CH · Switzerland

Affected area 6

CN · China

Affected area 7

ES · Spain

Affected area 8

GB · United Kingdom

Affected area 9

HK · Hong Kong

Affected area 10

KR · South Korea

Affected area 11

LT · Lithuania

Affected area 12

NZ · New Zealand

Affected area 13

SA · Saudi Arabia

Affected area 14

TW · Taiwan

Affected area 15

ZA · South Africa

Company & contacts

Brand

Manufacturer

Steris Corporation

Phone

Not provided

Brand: Steris. Contact: Steris Corporation.

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-1595-2023

Event ID

92044

Units affected

322

Status

active

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 24 May 2023. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button