Open official notice

Phasitron 5, IPV In-Line Valve, (Intrapulmonary Percussive Ventilation) REF: P5-TEE-20 (pack of 20) P5-TEE (individual) with In-Line Valve

Official notice

22 Oct 2025

RecallScope risk estimate:CriticalFDA

Phasitron 5, IPV In-Line Valve, (Intrapulmonary Percussive Ventilation) REF: P5-TEE-20 (pack of 20)/ P5-TEE (individual) with In-Line Valve Sentec Instructions for Use

What to do now

Stop using the affected device and follow the corrective action in the official FDA notice. Contact your healthcare provider or the manufacturer for next steps.

Units affected

59,160

Recall number

Z-0031-2026

Official notice

2025-10-22

The hazard

IFU update; in-line valve, with pressure relief seal may leak when used with ventilator; don't use in neonates/infants, with target tidal volumes less-than50mL or less-than10kg- at high risk of cardiopulmonary/neurological compromise due to unrecognized hypoventilation, updates for continued use in others before design update; leak may cause respiratory acidosis, hypoxemia, or respiratory failure.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Other

UDI-DI 00849436000723

Event Ids

97374

Recall Numbers

Z-0031-2026

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Phasitron 5, IPV In-Line Valve, (Intrapulmonary Percussive Ventilation) REF: P5-TEE-20 (pack of 20)/ P5-TEE (individual) with In-Line Valve Sentec Instructions for Use

Units affected

59,160

Hazard

IFU update; in-line valve, with pressure relief seal may leak when used with ventilator; don't use in neonates/infants, with target tidal volumes less-than50mL or less-than10kg- at high risk of cardiopulmonary/neurological compromise due to unrecognized hypoventilation, updates for continued use in others before design update; leak may cause respiratory acidosis, hypoxemia, or respiratory failure.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Affected area 2

AE · United Arab Emirates

Affected area 3

CA · Canada

Affected area 4

CH · Switzerland

Affected area 5

CL · Chile

Affected area 6

GB · United Kingdom

Affected area 7

IL · Israel

Affected area 8

JP · Japan

Affected area 9

RU · Russia

Affected area 10

TR · Turkey

Affected area 11

ZA · South Africa

Company & contacts

Manufacturer

Percussionaire Corporation

Phone

Not provided

Brand: Percussionaire. Contact: Percussionaire Corporation.

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class I

Recall details

Recall number

Z-0031-2026

Event ID

97374

Units affected

59,160

Status

active

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 22 Oct 2025. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button