An official Health Canada notice concerns Zeus® infinity® empowered. In the course of service activities, it was found that CO2 readings were slightly increased. In some cases, the ISO accuracy was no longer achieved. The tolerance check is intended to ensure that the zeus ie will measure all gases within the specified accuracy until the next annual check. Draeger has updated the instructions for use of the Zeus® ie anesthesia device regarding the test tolerance of the CO2 measurement. Recall start date: December 23, 2024 Published 2025-01-07.
What to do now
Contact the manufacturer if you require additional information.
Recall number
76804
Official notice
2025-01-07
The hazard
The hazard
In the course of service activities, it was found that CO2 readings were slightly increased. In some cases, the ISO accuracy was no longer achieved. The tolerance check is intended to ensure that the zeus ie will measure all gases within the specified accuracy until the next annual check. Draeger has updated the instructions for use of the Zeus® ie anesthesia device regarding the test tolerance of the CO2 measurement. Recall start date: December 23, 2024
How to check if you're affected
How to check if you're affected
Compare the VIN on your vehicle against the official list. If your VIN is listed, follow the official notice.
Recall Numbers
Checked manually this time?
The app watches your VIN and alerts you automatically next time — no searching, no checking back.
Turn on automatic VIN checksAffected products
Affected products
Product 1
Risk / hazard
Where it was sold
Where it was sold
The source does not list specific retailers. See the official notice for where affected vehicles were sold.
Location 1
Affected area 1
Company & contacts
Company & contacts
Brand
Manufacturer
Phone
Brand: Draegerwerk Kgaa, Zeus® Infinity® Empowered.
Source & status
Source & status
Source
Tier
Category
Classification
App estimate
Recall details
Recall details
Recall number
Classification
Source tier
Documents, notices & source
See the official notice and source documents for full details.
View full official noticeRelated recalls

Atlan a350

Combistar, safestar and twinstar filters hmes

Zeus infinity empowered (anesthesia workstation)

Ergostar cm 45 and ergostar cm 55

Vista 120 and vista 120s

Zeus infinity empowered

Ponta ceiling supply unit

Primus and zeus infinity aneasthetic workstations

Perseus a500

Atlan® a350

Filter hme twinstar plus

Carina ventilators
Weekly recall digest
Get next week's recalls before you need them.
One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.
The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.
This recall was published 7 Jan 2025. The next one won't wait for you to search.
RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.