No image published for this recall
Filter hme twinstar plus
Official notice
4 Aug 2023
Last checked
4 Aug 2023
An official Health Canada notice concerns Filter hme twinstar plus. Manufacturer's market surveillance activities revealed cases of increased inspiratory resistance resulting in insufficient ventilation, which may be related to formation of condensate in the filter housing. Device is intended exclusively for single patient use no longer than 24 hours. Excessive condensation can significantly shorten the period of use of the device and require the product to be replaced latest after 24 hours. Increased resistance carries the risk of desaturation for the patient. Should this hazardous situation encounter, consider the possibility that it may be related to the occlusion of the filter due to condensate. The current instructions for use contain warnings related to condensate. However, manufacturer has determined that the instructions can be further improved. Until the revised instructions for use are available customers may continue using the device as long as both airway pressure and volume are permanently monitored and alarm limits for each patient are suitably selected. Change of filter is required in the case of resistance increase. Recall start date: July 24, 2023 Published 2023-08-04.
What to do now
Contact the manufacturer if you require additional information.
Recall number
74068
Official notice
2023-08-04
The hazard
The hazard
Manufacturer's market surveillance activities revealed cases of increased inspiratory resistance resulting in insufficient ventilation, which may be related to formation of condensate in the filter housing. Device is intended exclusively for single patient use no longer than 24 hours. Excessive condensation can significantly shorten the period of use of the device and require the product to be replaced latest after 24 hours. Increased resistance carries the risk of desaturation for the patient. Should this hazardous situation encounter, consider the possibility that it may be related to the occlusion of the filter due to condensate. The current instructions for use contain warnings related to condensate. However, manufacturer has determined that the instructions can be further improved. Until the revised instructions for use are available customers may continue using the device as long as both airway pressure and volume are permanently monitored and alarm limits for each patient are suitably selected. Change of filter is required in the case of resistance increase. Recall start date: July 24, 2023
How to check if you're affected
How to check if you're affected
Compare the VIN on your vehicle against the official list. If your VIN is listed, follow the official notice.
Recall Numbers
Checked manually this time?
The app watches your VIN and alerts you automatically next time — no searching, no checking back.
Turn on automatic VIN checksAffected products
Affected products
Product 1
Risk / hazard
Where it was sold
Where it was sold
The source does not list specific retailers. See the official notice for where affected vehicles were sold.
Location 1
Affected area 1
Company & contacts
Company & contacts
Brand
Manufacturer
Phone
Brand: Draegerwerk Kgaa, Filter Hme Twinstar Plus.
Source & status
Source & status
Source
Tier
Category
Classification
App estimate
Recall details
Recall details
Recall number
Classification
Source tier
Documents, notices & source
See the official notice and source documents for full details.
View full official noticeRelated recalls

Atlan a350

Ergostar cm 45 and ergostar cm 55

Combistar, safestar and twinstar filters hmes

Zeus infinity empowered (anesthesia workstation)

Vista 120 and vista 120s

Zeus® infinity® empowered

Zeus infinity empowered

Ponta ceiling supply unit

Primus and zeus infinity aneasthetic workstations

Perseus a500

Atlan® a350

Carina ventilators
Weekly recall digest
Get next week's recalls before you need them.
One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.
The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.
This recall was published 4 Aug 2023. The next one won't wait for you to search.
RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.