
No image published for this recall
Atlan® a350
Official notice
7 Feb 2024
Last checked
7 Feb 2024
An official Health Canada notice concerns Atlan® a350. The recall is being initiated as the market surveillance activities revealed a few cases in which the internal backup battery failed spontaneously while the Atlan® is being operated without mains supply. This resulted in an unexpected shutdown of the device while it was running on batteries. The affected Atlans started with a battery status of 100%, but quickly shut down and did not trigger the alarm for low battery. A secondary acoustic alarm signal which is independent from mains and battery power was generated as specified. No patient consequences have been reported to Draeger so far. Recall start date: November 21, 2023 Published 2024-02-07.
What to do now
Contact the manufacturer if you require additional information.
Recall number
75032
Official notice
2024-02-07
The hazard
The hazard
The recall is being initiated as the market surveillance activities revealed a few cases in which the internal backup battery failed spontaneously while the Atlan® is being operated without mains supply. This resulted in an unexpected shutdown of the device while it was running on batteries. The affected Atlans started with a battery status of 100%, but quickly shut down and did not trigger the alarm for low battery. A secondary acoustic alarm signal which is independent from mains and battery power was generated as specified. No patient consequences have been reported to Draeger so far. Recall start date: November 21, 2023
How to check if you're affected
How to check if you're affected
Compare the VIN on your vehicle against the official list. If your VIN is listed, follow the official notice.
Recall Numbers
Checked manually this time?
The app watches your VIN and alerts you automatically next time — no searching, no checking back.
Turn on automatic VIN checksAffected products
Affected products
Product 1
Product 2
Risk / hazard
Where it was sold
Where it was sold
The source does not list specific retailers. See the official notice for where affected vehicles were sold.
Location 1
Affected area 1
Company & contacts
Company & contacts
Brand
Manufacturer
Phone
Brand: Draegerwerk Kgaa, Atlan® A350.
Source & status
Source & status
Source
Tier
Category
Classification
App estimate
Recall details
Recall details
Recall number
Classification
Source tier
Documents, notices & source
See the official notice and source documents for full details.
View full official noticeRelated recalls

Atlan a350

Zeus infinity empowered (anesthesia workstation)

Ergostar cm 45 and ergostar cm 55

Combistar, safestar and twinstar filters hmes

Vista 120 and vista 120s

Zeus® infinity® empowered

Zeus infinity empowered

Ponta ceiling supply unit

Perseus a500

Filter hme twinstar plus

Carina ventilators

Oxylog 3000 plus - basic unit
Weekly recall digest
Get next week's recalls before you need them.
One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.
The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.
This recall was published 7 Feb 2024. The next one won't wait for you to search.
RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.