Xps system-blades

Official notice

1 Sept 2021

RecallScope risk estimate:CriticalHealth Canada

An official Health Canada notice concerns Xps system-blades. Medtronic is voluntarily recalling specific lot numbers of sterile blade single-use accessories. Navigation blades are being used to aid in accuracy during the procedure, and if they wobble, that accuracy and precision is diminished. Medtronic has received complaints related to tip breaks and wobble/vibration in this subset of devices. Tip breaks or detachment of fragments may cause harm to the patient, cause lack of oxygenation or foreign body reaction if the fragment were to migrate into the patient's airway or be irretrievable. There is also a chance that the tip may impact critical structures such as nerves and arteries. Blade wobble or vibration may also lead to permanent damage if it causes the blade to impact critical structures. Recall start date: August 16, 2021 Published 2021-09-01.

What to do now

Contact the manufacturer if you require additional information.

Recall number

64448

Official notice

2021-09-01

The hazard

Medtronic is voluntarily recalling specific lot numbers of sterile blade single-use accessories. Navigation blades are being used to aid in accuracy during the procedure, and if they wobble, that accuracy and precision is diminished. Medtronic has received complaints related to tip breaks and wobble/vibration in this subset of devices. Tip breaks or detachment of fragments may cause harm to the patient, cause lack of oxygenation or foreign body reaction if the fragment were to migrate into the patient's airway or be irretrievable. There is also a chance that the tip may impact critical structures such as nerves and arteries. Blade wobble or vibration may also lead to permanent damage if it causes the blade to impact critical structures. Recall start date: August 16, 2021

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Recall Numbers

64448

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Affected products - Xps System-Blades, Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalog number - 1883516HRE 1882916 1884380HR 1883080HRE 1883506HRE 1883514 1883480HRE 1884013 1882936E 1884005HRE 1884030HRE 1884033HRE 1882902HRE 1882924HRE 1883502HRE 1883512HRE 1884031HRE 1884032 1882905HRE 1884020

Hazard

Medtronic is voluntarily recalling specific lot numbers of sterile blade single-use accessories. Navigation blades are being used to aid in accuracy during the procedure, and if they wobble, that accuracy and precision is diminished. Medtronic has received complaints related to tip breaks and wobble/vibration in this subset of devices. Tip breaks or detachment of fragments may cause harm to the patient, cause lack of oxygenation or foreign body reaction if the fragment were to migrate into the patient's airway or be irretrievable. There is also a chance that the tip may impact critical structures such as nerves and arteries. Blade wobble or vibration may also lead to permanent damage if it causes the blade to impact critical structures. Recall start date: August 16, 2021

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Medtronic Xomed, Xps System-blades.

Source & status

Tier

official

Category

Home & Consumer

Official classification

Type II

Recall details

Recall number

64448

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 1 Sept 2021. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button