Open official notice

VNS Therapy Programmer, Model 3000 used with Models 103, 104, 105, 106, or 8103 VNS Therapy generators

Official notice

1 Apr 2020

RecallScope risk estimate:CriticalFDA

An official FDA notice concerns VNS Therapy Programmer, Model 3000 used with Models 103, 104, 105, 106, or 8103 VNS Therapy generators. False positive warning may occur after: 1) VNS Generator interrogated at 0mA normal output current, 2) Generator programmed to non-0mA output current, 3) In-session re-interrogation performed. Users instructed to lower output current and widen pulse width. Only system diagnostic testing evaluates output current. Users may conclude device malfunction, could lead to medical/surgical intervention. Published 2020-04-01.

What to do now

Stop using the affected device and follow the corrective action in the official FDA notice. Contact your healthcare provider or the manufacturer for next steps.

Units affected

2,683

Recall number

Z-1563-2020

Official notice

2020-04-01

The hazard

False positive warning may occur after: 1) VNS Generator interrogated at 0mA normal output current, 2) Generator programmed to non-0mA output current, 3) In-session re-interrogation performed. Users instructed to lower output current and widen pulse width. Only system diagnostic testing evaluates output current. Users may conclude device malfunction, could lead to medical/surgical intervention.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

84723

Recall Numbers

Z-1563-2020

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

VNS Therapy Programmer, Model 3000 used with Models 103, 104, 105, 106, or 8103 VNS Therapy generators.

Units affected

2,683

Hazard

False positive warning may occur after: 1) VNS Generator interrogated at 0mA normal output current, 2) Generator programmed to non-0mA output current, 3) In-session re-interrogation performed. Users instructed to lower output current and widen pulse width. Only system diagnostic testing evaluates output current. Users may conclude device malfunction, could lead to medical/surgical intervention.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States · AK

Location 2

US · United States · AL

Location 3

US · United States · AR

Location 4

US · United States · AZ

Location 5

US · United States · CA

Location 6

US · United States · CO

Location 7

US · United States · CT

Location 8

US · United States · DC

Location 9

US · United States · DE

Location 10

US · United States · FL

Location 11

US · United States · GA

Location 12

US · United States · HI

Location 13

US · United States · IA

Location 14

US · United States · ID

Location 15

US · United States · IL

Location 16

US · United States · IN

Location 17

US · United States · KS

Location 18

US · United States · KY

Location 19

US · United States · LA

Location 20

US · United States · MA

Location 21

US · United States · MD

Location 22

US · United States · ME

Location 23

US · United States · MI

Location 24

US · United States · MN

Location 25

US · United States · MO

Location 26

US · United States · MS

Location 27

US · United States · MT

Location 28

US · United States · NC

Location 29

US · United States · ND

Location 30

US · United States · NE

Location 31

US · United States · NJ

Location 32

US · United States · NM

Location 33

US · United States · NV

Location 34

US · United States · NY

Location 35

US · United States · OH

Location 36

US · United States · OK

Location 37

US · United States · OR

Location 38

US · United States · PA

Location 39

US · United States · RI

Location 40

US · United States · SC

Location 41

US · United States · SD

Location 42

US · United States · TN

Location 43

US · United States · TX

Location 44

US · United States · UT

Location 45

US · United States · VA

Location 46

US · United States · VT

Location 47

US · United States · WA

Location 48

US · United States · WI

Location 49

US · United States · WV

Location 50

US · United States · WY

Affected area 1

US · United States

Affected area 2

CA · Canada

Affected area 3

IL · Israel

Company & contacts

Manufacturer

LivaNova USA Inc

Phone

Not provided

Brand: Livanova. Contact: LivaNova USA Inc.

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-1563-2020

Event ID

84723

Units affected

2,683

Status

active

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 1 Apr 2020. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button