Open official notice

VITROS Immunodiagnostic Products TSH Reagent Pack, REF Catalog Number Product Code 191 2997, Unique Device Identifier No. 10758750000227

Official notice

24 Aug 2016

RecallScope risk estimate:LowFDA

VITROS Immunodiagnostic Products TSH Reagent Pack, REF/Catalog Number/Product Code 191 2997, Unique Device Identifier No. 10758750000227, IVD --- Ortho-Clinical Diagnostics Pencoed Bridgend CF35 5PZ UK For the in vitro quantitative measurement of thyroid stimulating hormone (TSH) in human serum and plasma (EDTA or heparin) using the VITROS ECi/ECiQ Immunodiagnostic Systems, the VITROS 3600 Immunodiagnostic System and the VITROS 5600 Integrated System to aid in the differential diagnosis of thyroid disease.

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

2,883

Recall number

Z-2580-2016

Official notice

2016-08-24

The hazard

A field correction was issued by the firm for VITROS TSH reagent lot 5040 due to a higher than expected frequency of customer calibration curves falling outside of the calibration quality parameters and therefore unusable to process samples. Customers using VITROS TSH Lot 5040 will potentially experience an unexpected increase in the unsuccessful calibrations potentially causing a delay in testing and reporting results for VITROS TSH.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Lots

5040

Event Ids

74828

Recall Numbers

Z-2580-2016

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

VITROS Immunodiagnostic Products TSH Reagent Pack, REF/Catalog Number/Product Code 191 2997, Unique Device Identifier No. 10758750000227, IVD --- Ortho-Clinical Diagnostics Pencoed Bridgend CF35 5PZ UK For the in vitro quantitative measurement of thyroid stimulating hormone (TSH) in human serum and plasma (EDTA or heparin) using the VITROS ECi/ECiQ Immunodiagnostic Systems, the VITROS 3600 Immunodiagnostic System and the VITROS 5600 Integrated System to aid in the differential diagnosis of thyroid disease.

Units affected

2,883

Hazard

A field correction was issued by the firm for VITROS TSH reagent lot 5040 due to a higher than expected frequency of customer calibration curves falling outside of the calibration quality parameters and therefore unusable to process samples. Customers using VITROS TSH Lot 5040 will potentially experience an unexpected increase in the unsuccessful calibrations potentially causing a delay in testing and reporting results for VITROS TSH.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States · AR

Location 2

US · United States · GA

Location 3

US · United States · KY

Location 4

US · United States · MN

Location 5

US · United States · OK

Location 6

US · United States · SD

Location 7

US · United States · TN

Location 8

US · United States · VA

Location 9

US · United States · WV

Affected area 1

US · United States

Affected area 2

CA · Canada

Company & contacts

Manufacturer

ORTHO-CLINICAL DIAGNOSTICS

Phone

Not provided

Brand: Ortho-clinical Diagnostics. Contact: ORTHO-CLINICAL DIAGNOSTICS.

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class III

Recall details

Recall number

Z-2580-2016

Event ID

74828

Units affected

2,883

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 24 Aug 2016. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button