No image published for this recall

No image published for this recall

Vitek 2 system - susceptibility test cards

Official notice

2 Apr 2024

Last checked

2 Apr 2024

Medium riskHealth Canada

An official Health Canada notice concerns Vitek 2 system - susceptibility test cards. The St. Louis manufacturing site discovered a calculation error in the master formulation for ceftriaxone (cro02n) used in specific card lots. The error was related to the ceftriaxone concentration used in two wells of the test card. Recall start date: March 15, 2024 Published 2024-04-02.

What to do now

Contact the manufacturer if you require additional information.

Recall number

75312

Official notice

2024-04-02

The hazard

The St. Louis manufacturing site discovered a calculation error in the master formulation for ceftriaxone (cro02n) used in specific card lots. The error was related to the ceftriaxone concentration used in two wells of the test card. Recall start date: March 15, 2024

How to check if you're affected

Compare the product name and identifiers below with the item you own. Follow the official notice if they match.

Recall Numbers

75312

Checked manually this time?

The app alerts you when a new official recall is published — no searching, no checking back.

Turn on recall alerts

Affected products

Product 1

Affected products - Vitek 2 System - Susceptibility Test Cards, Lot or serial number - 1402761504 1412754504, Model or catalogue number - 423340 423341

Risk / hazard

The St. Louis manufacturing site discovered a calculation error in the master formulation for ceftriaxone (cro02n) used in specific card lots. The error was related to the ceftriaxone concentration used in two wells of the test card. Recall start date: March 15, 2024

Where it was sold

The source does not list specific retailers. See the official notice for where affected products were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Biomerieux, Vitek 2 System - Susceptibility Test Cards.

Source & status

Source

Health Canada

Tier

official

Category

HOME

Classification

Type II

App estimate

medium

Recall details

Recall number

75312

Classification

Type II

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 2 Apr 2024. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.