Open official notice

Trinity biotech oxalate reagent kit

Official notice

17 Sept 2026

Last checked

13 Nov 2026

RecallScope risk estimate:MediumHealth Canada

An official Health Canada notice concerns Trinity biotech oxalate reagent kit. Trinity Biotech is recalling Oxalate Reagent B, Product Code 591-2, Lot 2500711 due to performance variability resulting in negative bias and under-recovery of oxalate values, potentially causing falsely low patient results. Users should discontinue use, quarantine affected inventory, and follow recall instructions. Published 2026-09-17.

What to do now

Contact the manufacturer if you require additional information.

Recall number

82642

Official notice

2026-09-17

The hazard

Trinity Biotech is recalling Oxalate Reagent B, Product Code 591-2, Lot 2500711 due to performance variability resulting in negative bias and under-recovery of oxalate values, potentially causing falsely low patient results. Users should discontinue use, quarantine affected inventory, and follow recall instructions.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Lots

2500711

Skus

591-2

Other

RA RA-82642

Model numbers

591-2

Recall Numbers

82642

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Lot or serial number - 2500711, Model or catalogue number - 591-2

Hazard

Trinity Biotech is recalling Oxalate Reagent B, Product Code 591-2, Lot 2500711 due to performance variability resulting in negative bias and under-recovery of oxalate values, potentially causing falsely low patient results. Users should discontinue use, quarantine affected inventory, and follow recall instructions.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Trinity Biotech.

Source & status

Tier

beta

Category

Home & Consumer

Official classification

Type II

Recall details

Recall number

82642

Source tier

beta

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.